Welcome Customer !

Membership

Help

Anhui Chunhui Instrument Cable Group Co., Ltd
Custom manufacturer

Main Products:

ybzhan>Download Materials>Integrated vibration transmitter
Download

Anhui Chunhui Instrument Cable Group Co., Ltd

  • E-mail

    zy244295446@qq.com

  • Phone

    18205507022

  • Address

    No.1 Jing'er Road, Xicheng District, Tianchang City, Anhui Province

Contact Now
Integrated vibration transmitter
Date: 2022-04-01Read: 8

1. Introduction

The integrated vibration transmitter integrates piezoelectric sensors and precision measurement circuits together to form a high-precision vibration measurement system. The transmitter can be directly connectedDCS, PLC or other equipment are ideal choices for measuring the vibration of factory equipment such as fans, motors, pumps, etc.

2. Technical parameters

2.1Frequency response:5 ~ 1000Hz

2.2 Measurement Range:0~20.0mm/s

2.3 Output current: 4~20mA

2.4 Output impedance:≤500Ω

2.5 Working voltage: DC24V

2.6 Wiring method: 2-wire system

2.7Maximum acceleration:10g

2.8Measurement direction:Vertical or horizontal

2.9Usage environment: Warmdegree-40℃~85℃relative humidity≤90%

2.10Overall dimensions:φ24×60mm

2.11 Installation thread: M10 × 1.5 × 10mm (depth) or magnetic seat

2.12heavyQuantity: Approximately100g

3. An pretend

3.1 Installation position: Install vertically or horizontally on the measured vibration point, and place the sensor at the bottomMFix the 10 × 1.5 × 10 screws on the tested housing, and then tighten the sensor on top.

3.2 Installation dimensions and specifications: Refer to Figure 1. If the installation position of the transmitter is subject to high temperature steam erosion, protective measures need to be taken to reduce the ambient temperature of the transmitter. Generally, no protection is required.

3.3 Wiring instructions: Connect the red wireThe positive terminal of the DC24V power supply is connected to the negative terminal of the DC24V power supply through a black wire, and the red and black wires are also 4~20mA output lead.The connecting wire requires good insulation performance, and adoptstwoCore shielded cable.