Anhui Tiankang Group SBWR Thermocouple Temperature Transmitter Product Introduction
Date: 2025-06-25Read: 10

Anhui Tiankang Group SBWR thermocouple temperature transmitter
1、 Overview:
The SBWR thermocouple temperature transmitter and SBWZ thermistor temperature transmitter are on-site installed temperature transmission units in the DDZ-S series instruments. It adopts a two-wire transmission method (two wires as power input and a common transmission line for signal output). The transmitter can be installed in the junction box of the thermocouple and thermistor to form an integrated structure by converting the thermocouple and thermistor signals into input electrical signals, measured temperatures, or linear 4-20mA output signals. As a new generation of temperature measuring instrument, it can be widely used in industrial sectors such as metallurgy, petrochemicals, power, light industry, textiles, food, national defense, and scientific research.
2、 Characteristics of temperature transmitter:
1. Adopting an epoxy resin sealing structure, it is earthquake resistant, temperature resistant, and suitable for installation and use in harsh on-site environments.
2. Installed on-site in the junction box of thermal resistors and thermocouples, it directly outputs 4-20mA, which not only saves expensive compensation wire costs, but also improves the anti-interference ability during long-distance signal transmission.
3. High precision, low power consumption, wide temperature range for use, stable and reliable operation.
4. The range is adjustable and has a linearization correction function. The thermocouple temperature transmitter has a cold end automatic compensation function. Widely applicable, it can form an integrated on-site installation structure with thermocouples and thermistors, and can also be installed as a functional module in testing equipment.
3、 Main technical parameters:
1. Input: Thermal resistance Pt100, Cu50, Cu100
Thermocouples K, E, S, B, T, J, N
2. Output: Within the range of measurement, outputting a 4-20mA DC signal can be linear with the output resistance signal of the resistance thermometer and linear with the input temperature signal of the resistance thermometer; It can be linear with the millivolt signal input by the thermocouple, or linear with the temperature signal input by the thermocouple thermometer.
3. Basic error: ± 0.2%, ± 0.5%
4. Transmission method: two-wire system
5. The working power supply voltage of the transmitter is 12V, 35V, and the rated working voltage is 24V.


温度变送器
6. Load: The maximum load resistance is calculated according to the following formula
RL(max)=50×(Vmm-12)
At 24V, the load resistance can be selected within the range of 0-600 Ω, with a rated load of 250 Ω.
Note: The range adjustable transmitter requires repeated debugging of zero and full scale when changing the range; The thermocouple type transmitter needs to be preheated for 30 minutes before debugging.
7. Environmental temperature impact ≤ 0.05% 1 ℃
8. Normal working environment:
a、 Environmental temperature -25 ℃~+80 ℃
b、 Relative humidity 5% to 95%
c、 Mechanical vibration f ≤ 55Hz, amplitude < 0.15mm
4、 Model and category:

model Sensor indexing number *Large scale schedule regulations Specify the range of small values (difference between upper and lower limits)
SBWR-2160 E 0~800℃ 300℃
SBWR-2260 K 0~1300℃ 400℃
SBWR-2360 S 0~1600℃ 600℃
SBWR-2460 B 0~1800℃ 800℃
SBWR-2560 T 0~400℃ 200℃
SBWR-2660 J 0~1200℃ 300℃
SBWR-2760 N 0~1200℃ 400℃
SBWR-4160 E 0~800℃ 300℃
SBWR-4260 K 0~1300℃ 400℃
SBWR-4360 S 0~1600℃ 600℃
SBWR-4460 B 0~1800℃ 800℃
SBWR-4560 T 0~400℃ 200℃
SBWR-4660 J 0~1200℃ 300℃
SBWR-4760 N 0~1200 400℃
SBWZ-2160 Cu50 -50~150℃ 50℃
SBWZ-2260 Cu100 -50~150℃ 50℃
SBWZ-2480 Pt100 0~600℃ 50℃
SBWZ-2460 Pt100 -100~200℃ 50℃
SBWZ-4460 Pt100 -200~600℃ 50℃

5、 Main technical indicators:

Allowable deviation of thermocouple electromotive force

Thermocouple category graduation mark level Operating temperature range ℃ Allowable deviation Δ t ℃
Nickel chromium copper nickel E Ⅱ 0~600 ± 2.5 ℃ or 0.75% t
Nickel-chromium-nickel-silicon K Ⅱ 0~800 ± 2.5 ℃ or 0.75% t
Platinum rhodium 10 Platinum rhodium 5 S Ⅱ

0~600

600~1300

± 1.5 ℃ or 0.25% t
Platinum rhodium 30- Platinum rhodium B Ⅱ

600~800

800~1600

± 4 ℃ or 0.5% t
Copper Copper Nickel (Copper Copper) T Ⅱ -40~+350 ± 1 ℃ or 0.75% t
Nickel chromium silicon nickel silicon N Ⅱ -40~1200 ± 2.5 ℃ or 0.75% t