Instrument oxygen content analyzer data
Date: 2022-02-18Read: 12

NameNew zirconia oxygen analyzer

Model: HKZO402

Overview:

Our factory has developed a new generation of miniaturized zirconia analyzer based on the original technology, which uses high-precision electronic processors as the core module and solid zirconia electrolyte as the detection element to detect trace elements in mixed gases.

New zirconia oxygen analyzerThe functional characteristics,

The following measures were adopted during the design process: over limit sound and light alarm and signal output/upper and lower limit control points for the oxygen content of the gas to be tested can be set/gas sample purification pretreatment device. Reducing the failure rate during signal transmission while improving measurement accuracy. In addition, the sealed heating device and secondary over temperature protection function extend the service life of the heating furnace and improve the safety and reliability of the instrument. The new HKZO402 zirconia oxygen analyzer features a lightweight and exquisite appearance, small size, and small space occupation. There are two types: desktop and embedded/disc, which can be selected by users according to the on-site situation.

New zirconia oxygen analyzerPerformance parameters of

signal output

LED digital display, 0-10mA, 4-20mA, divided into six levels; The signal output automatically switches between different levels, and 1-5V output can be added according to demand.
Range gear: I 、 0.1-10 μ mol / mol(PPM) , II 、 10-100 μ mol / mol(PPM) ,
III 、 100-1000 μ mol / mol(PPM) , IV 、 0.1%-1%
V 、 1%-10% ,   VI 、 10%-

ambient temperature

0-40 ℃, P<100 μ mol / mol(PPM) ± 7.5 μ mol / mol(PPM) , P>100 μ mol / mol(PPM) ± 4.5 %;

measurement range

0.1 μ mol / mol(PPM)-100 % O 2

Basic error

P<100 μ mol / mol(PPM) ± 5 μ mol / mol(PPM) , P>100 μ mol / mol(PPM) ± 3 %

repeatability error

P<100 μ mol / mol(PPM) ± 2.5 μ mol / mol(PPM) , P>100 μ mol / mol(PPM) ± 1.5 %

Zero point and range drift

P<100 μ mol / mol(PPM) ± 5 μ mol / mol(PPM) / 24h ,P>100 μ mol / mol(PPM) ± 3 %/ 24h

Impact error

Voltage fluctuation: 220V ± 10%, P<100 μ mol/mol (PPM) ± 2.5 μ mol/mol (PPM), P>100 μ mol/mol (PPM) ± 1.5%

Frequency fluctuation: 50Hz ± 10%, P<100 μ mol/mol (PPM) ± 2.5 μ mol/mol (PPM), P>100 μ mol/mol (PPM) ± 1.5

Flow fluctuation: P<10 μ mol/mol (PPM) 200ml/min ± 5ml/min<± 5ppm
P<100 μ mol / mol(PPM) 200ml / min ± 10ml / min < ± 2 5ppm
P<100 μ mol / mol(PPM) 200ml / min ± 50ml / min< ± 3 %

time constant

<10s

Temperature control accuracy

750 ℃± 3 ℃

power consumption

50w 、zui 大 200W

warm-up time

1 h

appearance

Dimensions: 300 × 140 × 350mm (width × height × depth); (Type 401)
310 × 140 × 300mm (width × height × depth); (Type 402)
Opening: 312 ± 1 × 142 ± 1 mm (width × height)

Weight: 7.3kg

Output terminal wiring diagram: