Dielectric constant dielectric loss tester Impedance measuring instrumentModel: DP-ZJD-E
1Dielectric constant dielectric loss tester Impedance measuring instrumentBrief Introduction
The dielectric constant tester is a multifunctional component tester used for testing various electronic components. It adopts7-inch 800x480 color TFT LCD display, allowing users to intuitively read test results. It is an impedance measuring instrument with a resolution of 6 bits, which can meet the requirements of production line inspection and laboratory measurement.
IIDielectric constant dielectric loss tester Impedance measuring instrumentPerformance Characteristics
Four parameter display, any combination of parameters, up to12 measurement parameters;
7-inch 16:9 TFT LCD display with a resolution of 800x480 RGB;
0.05% basic measurement accuracy, fast testing speed (200 times/second);
Test frequency of 20Hz-1MHz, resolution of 10mHz;
GundamMeasurement speed of 200 times/s;
Automatic level control for testing voltage or currentALC function;
Four different signal source output impedances: 30 Ω, 100 Ω, 50 Ω, and 10/CC;
Built in comparator,10 level sorting and counting functions;
Frequency response curve plotting and analysis function;
10 point list scanning function;
internal-5V~+5V DC bias voltage source;
optional± 10V (± 100mA) and 1A internal DC bias source;
internalSave or call 100 sets of configuration files, and save or call 500 sets of test files on a USB drive;
can pass throughUSB HOST is used to upgrade and update instrument software versions;
The USB flash drive supports FAT16 and FAT32 file systems;
standard configurationRS232C、HANDLER、USB HOST, Earphone interface, optional GPIB and LAN interfaces;
3、 Technical indicators
Operating temperature and humidity0℃-40℃ , 90%RH
Power requirements voltage99V-121V 198V-242V
frequency47.5Hz-63Hz
power consumption≤80VA
Volume(W*H*D) 398mm×129mm×370mm
weightAbout 9kg
4、 Technical parameters
Measurement function test parameters|Z|, |Y| , C, L,X,B,R,G,D,Q, θ , DCR
Basic measurement accuracy0.05%
equivalent circuit Series connection, parallel connection
Mathematical functions Absolute value deviation, percentage deviation
Range method Automatic, hold, manual selection
Trigger method Internal, manual, external, bus
Measuring speed(≥ 1kHz) Fast: 75 times/second (fast)
Medium speed:50 times/second
Slow speed:5 times/second
average frequency1—255
delay time0—60s, Step by 1ms
Calibration functionOpen circuit/short circuit/load
Test end configuration Five ends
List scanning10 point list scanning function
display modeDirect reading, Δ, Δ%, V/I (measured voltage/current monitoring)
monitor6-digit resolution, 800 × 480 RGB 7-inch 16:9 TFT LCD display
test signal Test signal frequency20Hz-1MHz, 10mHz resolution
output impedance30 Ω, 100 Ω, 50 Ω, 10/CC can be selected
Test signal levelNormal: 5 mV to 2V
Accuracy:10%, 1 mV step
Constant voltage level:10 mV -1 V accuracy: 5%, 1mV step
display range|Z|, R, X 0.01m Ω — 99.9999 MΩ
DCR 1 mΩ — 99.9999 MΩ
|Y|, G, B 0.00001µS — 99.9999 S
C 0.00001 pF — 9.9999 F
L 0.00001 µH — 9999.99 H
D 0.00001 — 9.9999
Q 0.00001 — 9999.9
θ (DEG) -179.999 °— 179.999 °
θ (RAD) -3.14159 — 3.14159
other Comparator functionTen levels: (nine levels are qualified, one level is unqualified), and there is also an auxiliary level AUX
存储器100 sets of internal instrument settings for storage/retrieval
interfaceRS-232C, Handler, GPIB (optional), LAN (optional), USBTMC,USBCDC