Flame calibration device for vertical combustion instrument of single wire and cable
Date: 2018-05-08Read: 11

Single wire and cable vertical combustion instrument (quality inspection/testing unit model)
Flame calibration device for vertical combustion instrument of single wire and cableThis single wire and cable vertical combustion instrument is suitable for quality inspection/testing units and wire and cable production enterprises for acceptance or qualification review.
1、 Scope of use
The single wire and cable vertical combustion tester is mainly used to evaluate the combustibility of single wire and cable with a conductor diameter greater than 8mm (cross-sectional area greater than 0.5mm2) or less than 8mm (cross-sectional area less than 0.5mm2). Suitable for research, production, and quality inspection departments of electrical and electronic products and their component parts, such as lighting equipment, low-voltage appliances, household appliances, machine tool appliances, motors, power tools, electronic instruments, electrical instruments, electrical connectors, and accessories. This product adopts the recommended product structure for the air mixing structure in Appendix B of GB18380.12-2008, and the details also meet the acceptance requirements of the quality inspection unit.


2、 Product functional features
The vertical combustion test of a single wire and cable is conducted by using a Bunsen burner of a specified size and a specific gas source (propane) to ignite the specimen in a vertical state at a certain flame and flame angle. The combustibility and ignition hazard of the specimen are evaluated based on the duration and length of ignition and combustion.
3、 Main technical parameters
1. Seven inch full-color programmable controller touch screen+PLC control, realizing multifunctional control/detection/calculation/data display
2. High degree of automation: automatic recording of test time, automatic display of test results, automatic timing, automatic ignition, automatic retraction of the burner after flame application, and the option to turn off the gas
3. When starting, you can choose whether to automatically ignite; Start, stop, gas, timer, ignition, save, save, lighting, and exhaust are all controlled through a touch screen, and the experiment can be completed with a simple touch of your fingertips
4. The timing button, in conjunction with the PLC, automatically records and stores the combustion time consumed
5. The test time system automatically records and generates test results
6. Burner: Inner diameter Φ 9.5mm ± 0.5mm
7. Test inclination angle: The 45 ° style crossbeam can move forward and backward, and the position of the lamp can be adjusted according to the thickness of the style;
8. Flame: adjustable from 20mm ± 2mm to 190mm ± 1mm
9. Flame application time: adjustable from 0-99.9s ± 0.1s
10. Combustion gas: 95% propane gas (usually replaced by liquefied petroleum gas)
11. Gas: high-purity propane gas or liquefied petroleum gas (to be provided by the user) is used;
12. Tested wire and cable length: 600 ± 25mm;
13. Outer diameter of tested wires and cables: conductor diameter greater than 0.8mm or cross-sectional area greater than 0.5mm2
14. This device should be tested in an environment with no wind and a temperature of (23 ± 10) ℃
15. Standard power of combustion blowtorch: 1KW;
16. Power supply voltage: 220V/50Hz
17. Gas flow rate: 0-1L/min
18. Air flow rate: 1.5-15L/min
19. Box material: Steel plate electrostatic spraying (stainless steel material box can be used)
20. Equipment dimensions: length 60 cm, width 45 cm, height 156 cm


4、 Design standards
GB/T18380.1-2008、GB/T18380.11-2008、GB/T18380.12-2008、
GB/T18380.21-2008、GB/T18380.22-2008、GB/T5169.14-2007
5、 Applicable standards
IEC60332、GB/T19666-2005、GB12972、GB13033、MT818、MT386、ISO5656
Another standard recommends two flame calibration methods:
(1) Flame inner flame 50-60mm, outer flame 170-190mm
(2) The time for heating 10g copper head from 100 ℃ to 700 ℃ is 45 ± 5 seconds
If the user uses the second method to verify the flame, a flame calibration device needs to be selected. The flame calibration method is not currently considered for early procurement, and can be added later if necessary.
The DJC-1 type single wire and cable vertical combustion testing machine meets the standard:
GB/T18380.11-2008 Single Insulated Wire and Cable Flame Vertical Spread Test Device
GB/T18380.12-2008 Single insulated wire and cable vertical flame propagation test 1kW premixed flame test method
GB/T18380.21-2008 Single Insulated Thin Wire and Cable Flame Vertical Spread Test Device
GB/T18380.22-2008 Single insulated thin wire and cable flame vertical spread test diffusion flame test method
IEC60332-1-2004
GB/T5169.14-2007 Fire hazard testing for electrical and electronic products Part 14: Test flame 1kW nominal premixed flame equipment, confirmation test methods and guidelines
GB/T19666-2005 General rules for flame-retardant and fire-resistant wires and cables
GB12972 Mining rubber sheathed flexible cable
GB13033 Mineral Insulated Cables with Rated Voltage of 750V and Below
MT818、MT386、ISO5656