TS-2-002Weighing amplifier
1、 Product Features
This product is launched according to the needs of on-site industrial occasions,This amplifier can1~4The road weighing force sensormVThe signal is amplified and transmitted to the secondary instrument and central control room for display and monitoring.
Die cast aluminum shell, sturdy and durable. Equipped with voltage reverse protection. High precision operational amplifier circuit, with high precision and good stability. Low temperature drift and good anti-interference performance. The sensitivity input range is wide.
2、 Main technical indicators
1Input1~4group0-24mVThe weighing sensor.
2、Output signal type:0-10mA 4-20mA 0-10V 0-5V 1-5V 0-100mV(Customer Signal Type)
3Amplifier power supply:DC24V DC12V(Customer power supply).
4Working temperature:-20~80℃
5Working humidity:<90%.
6Accuracy: superior to0.1%FS.
7Shell protection:IP65.
3、 Sensor debugging
1Sensor adjustment
Firstly, load the standard weight onto each sensor and measure the output value of each sensor. Calculate the average of all measured values and adjust accordinglyRS1~RS4Use the potentiometer until the adjusted value reaches the standard measurement average. If it cannot be adjusted at once, the previous operation can be repeated.
2Zero point adjustment of transmitter output
Firstly, place the system in an unloaded state and adjust the marking (zero point)ZER0Use a potentiometer to adjust the deviation of the zero point. Adjust the indication to the zero point indication.
3Gain adjustment
Load the system with a certain weight value (recommended range)60%~)Calculate the output value based on the proportional relationship between the loaded weight and the large weight value, and adjust it accordingly(full scaleFULLThe potentiometer adjusts the reading to the loading ratio reading.
4Re examination
Place the system in an unloaded state and measure the zero point output: load a certain weight value and measure whether the output meets the expected amplification factor. If not, please repeat the steps2、3Adjust again.
4、 Installation dimensions
5、 Wiring instructions
Connect sensor end (five core):1Positive power supply+,4negative power supply- 2Signal positive+,3Signal negative-.
Power output terminal (four core):1Positive power supply+red thread, 2Signal output positive+Green line, 3~4Public negative-Blue line and black line.
6、 System connection