Standardized use and regular maintenance of full temperature oscillation incubator
Date: 2025-11-14Read: 10

The full temperature oscillation incubator can be used to simulate microbial metabolism research under different environmental conditions by adjusting temperature and oscillation frequency, enabling rapid temperature switching and liquid mixing of the equipment. It can enhance the contact between the culture medium and cells and promote cell proliferation. Optimizing the cultivation conditions significantly improved experimental efficiency and data reliability.

The stable operation of the full temperature oscillation incubator relies on the dual guarantee of "standardized use+regular maintenance". Common usage methods and maintenance guidelines, Changzhou Gaode Instrument Manufacturing will introduce you to help you efficiently manage equipment.

1、 Power check before startup: Use a 220V ± 10%, 50Hz power supply and ensure good grounding. The power carrying capacity of the socket should be ≥ 1kW to avoid sharing with other high-power devices. Environmental requirements: The equipment should be placed on a well ventilated workbench away from direct sunlight, with a distance of at least 20cm between the back and the wall to facilitate heat dissipation. It is recommended to control the ambient temperature between 10-30 ℃ and humidity ≤ 85%.

2、 Setting and operation of oscillation and temperature parameters:

1. Setting of oscillation parameters: After turning on the power switch, set the speed (60-260rpm) through the digital speedometer. Using medium speed oscillation can effectively extend the service life of the equipment. The equipment is equipped with a three-dimensional integrated three wheel drive system, which runs smoothly and can maintain high stability even after long-term use. This design effectively reduces mechanical wear, extends equipment life, and reduces the risk of sample splashing during oscillation.

2. Temperature cultivation setting: Press the "SET" button to enter the temperature setting mode, and use the "▲/▼" buttons to adjust to the target temperature (range 0-60.0 ℃), which can be set arbitrarily to meet diverse needs from low-temperature storage to high-temperature fermentation.

3. The timing function supports timed switching operation from 0-9999m/h or "ON" normally open mode. After startup, it automatically maintains a constant temperature, speed, and timing.

4. Sample placement: Place the culture bottles symmetrically on the tray, and it is recommended to place them evenly to avoid oscillation and imbalance. Triangle bottles or test tubes should be placed stably and firmly to prevent tipping.

5. Operation monitoring: After startup, observe the temperature stability and oscillation uniformity, and check the internal status through the lighting. Avoid frequent door opening to prevent temperature fluctuations.

3、 Shutdown process: After the experiment is completed, press the "STOP" button to stop oscillation, remove the sample, and turn off the power. When not in use for a long time, unplug the power cord and clean the internal cavity.

4、 Common troubleshooting:

Abnormal temperature: Press and hold the "SET" button for 10 seconds to restore the default settings and recalibrate the temperature.

Unstable oscillation: Check the balance of the tray, tighten the screws or replace the belt.

Compressor noise: Avoid continuous start-up, ensure good heat dissipation, and contact maintenance if necessary.