Liquid: When the density change is ≤ 5%, the error can be ignored; If the density change is greater than 10% (such as high and low temperature oil products), recalibration or compensation is required.
Gas: Density is strongly correlated with temperature and pressure (ρ∝ P/T), and accuracy needs to be corrected through temperature and pressure compensation.
example
Measure natural gas (standard density 1.2 kg/m ³), if the operating pressure is 0.5 MPa and the temperature is 300 K, the actual density is:
When not compensated, density deviation can lead to flow errors of over ± 10%.
Insufficient straight pipe section:
When the upstream straight pipe section is less than 5D, turbulence causes the float to shake, and the error may increase to ± 3% -5%.
Solution: Install a rectifier or choose an anti turbulence flow meter.
Tilt installation:
When the vertical installation tilt is greater than 2 °, the friction between the float and the cone tube changes, and the error increases by ± 1% -2%.
Solution: Use a horizontally installed flow meter or adjust the installation angle.
Pulsating flow:
Under pulsating flow conditions such as compressor outlet and pump rear, float vibration causes reading fluctuations with an error of up to ± 5% -10%.
Solution: Install buffer tanks or dampers, or choose intelligent flow meters with digital filtering function.
Temperature impact:
The cone tube and float of the metal tube float flowmeter have different coefficients of thermal expansion. When the temperature changes by more than 50 ℃, the range may shift by ± 1% -2%.
Solution: Choose a flow meter with temperature compensation or perform regular calibration.
Pressure shock:
When high-pressure gas is rapidly released, the float may impact the top of the cone tube due to inertia, resulting in instantaneous readings exceeding the tolerance or damage.
Solution: Install a voltage regulator valve or current limiting device.
Human error in reading:
The glass tube float flowmeter requires manual reading, and visual deviation may result in an error of ± 1% -2%.
Solution: Choose an intelligent flowmeter with 4-20mA output or install a photoelectric reading device.