Introduction to the use precautions of carbon dioxide incubator on the main platform
Date: 2016-01-04Read: 11
1CO2 incubatorBefore injecting water, do not turn on the power switch, otherwise it will damage the heating element.
2CO2 incubatorAfter running for several months, the water in the water tank may decrease due to evaporation. When the low water level indicator light is on, water should be replenished. First, open the overflow pipe and use a funnel to connect the rubber tube to add water from the injection hole to turn off the low water level indicator light. Then, measure the amount of water added and block the overflow hole.
3CO2 incubatorA high-precision constant temperature incubator can be used, but the CO2 control system needs to be turned off.
4. Because the CO2 sensor is calibrated under saturated humidity, the humidification tray must always be filled with sterile water.
5. When the displayed temperature exceeds the set temperature by 1 ℃, the over temperature alarm indicator light will turn on and emit a sharp alarm sound. At this time, the power should be turned off for 30 minutes; If the power (temperature control) switch is turned on again but still exceeds the temperature limit, the power should be turned off and maintenance personnel should be reported.
6. When the pressure of the steel cylinder is lower than 0.2MPa, the cylinder should be replaced.
7. Try to minimize the time it takes to open the glass door.
8. IfCO2 incubatorIf not in use for a long time, the moisture in the studio must be removed before closing, and the glass door must be opened for ventilation for 24 hours before closing.
9. When cleaning the incubator workshop, do not collide with components such as sensors, mixing motors, and wind turbines.
10. When disassembling and assembling the bracket cover in the workshop, a random wrench must be used and excessive force must not be applied.
11. MovingCO2 incubatorThe water inside the box must be drained before. When draining, tightly fit the rubber hose over the water outlet hole so that the pipe mouth is lower than the instrument, gently suck in a mouthful, lower the water pipe, and the water will be siphoned out.
12. MovingCO2 incubatorPreviously, the shelves and humidification tray in the studio should be taken out to prevent collision and damage to the glass door.
13. MovingCO2 incubatorWhen doing so, do not invert and do not lift the box door to prevent it from deforming.
Dongguan Zhengtai Testing Instrument Co., LtdContinuously exploring and innovating, satisfying. With the rapid changes in the market and the diverse needs of our customers, we have devoted ourselves to exploring and researching, introducing advanced Japanese technology. This not only improves the performance and reliability of our products, but also effectively reduces manufacturing costs and feeds back profits to our customers, enabling them to have high-quality and reasonably priced products. Welcome all users to come and negotiate.