Methods for handling damaged accessories in cold and hot shock tests
1、 Core component damage handling plan
Temperature sensor failure
Diagnosis: Use a multimeter to measure the resistance value. If the deviation from the standard value exceeds 5%, it needs to be replaced
Replacement operation: Disassemble and install after power failure, calibrate PID parameters after replacement to ensure temperature control accuracy
The aging of the sealing strip leads to temperature leakage
Temporary handling: Clean the sealing groove and apply silicone grease to temporarily enhance the sealingSolution: Replace the entire sealing strip and prioritize using silicone rubber material with a temperature resistance range of -70 ℃~150 ℃
Compressor malfunction
Refrigerant leakage: When detecting abnormal pressure gauge values, repair the leakage point first and then charge the refrigerant according to the standard
Mechanical failure: When the operating current exceeds the tolerance of ± 15% or there is abnormal vibration, it is necessary to perform maintenance on the bearings/valve plates
2、 Response to auxiliary system failures
Abnormal wind circulation system
Fan shutdown: Remove dust from fan blades and test motor winding resistance
Air duct blockage: Remove the air guide plate to clean foreign objects, optimize sample placement to avoid obstructing airflow
Humidification system malfunction
Humidity out of control: Check if the wet bulb gauze has hardened (replace monthly)
Detecting the sensitivity of the humidification boiler water level controller
Control system alarm
PID parameter misalignment: Reset the proportional band/integral time parameters according to the equipment manual
Overtemperature protection triggered incorrectly: verify the on/off status of the solid-state relay
Monthly maintenance
Clean the dust on the condenser fins (blown with compressed air)
Tighten the electrical terminals to prevent poor contact.
Annual deep maintenance
Replace all seals and sensors (whether damaged or not)
Vacuumize the refrigeration system and refill it with refrigerant