Malfunctions and solutions encountered during the operation of salt spray corrosion test chamber
Date: 2022-11-17Read: 12

Due to the long-term salt spray aging test conducted by the salt spray testing machine, the harsh environment can easily cause various malfunctions of the salt spray testing machine. In order to quickly solve the problems and continue to conduct salt spray aging tests smoothly, the following provides a detailed introduction to some common malfunctions and their solutions.

The laboratory cannot rise to the set temperature

Reason for malfunction:

1. The temperature of the laboratory temperature controller is set too low.

2. The safety protection switch in the laboratory is set too low.

3. Heating system malfunction.

4. Electromagnetic relay malfunction.

5. Controller malfunction.

Handling method:

1. Set the temperature controller to the desired temperature.

2. Set the safety protection switch to the desired temperature.

3.4.5. Notify the repair company to come for on-site repairs.

The temperature of the saturated tank cannot rise to the set temperature

Reason for malfunction:

1. The temperature of the saturation tank temperature controller is too low.

2. The saturation bucket safety protection switch is set too low.

3. Heating system malfunction.

4. Electromagnetic relay malfunction.

5. Controller malfunction.

Handling method:

1. Set the temperature controller to the desired temperature.

2. Set the safety protection switch to the desired temperature.