How should the DI5 alarm of the constant temperature and humidity test chamber be handled?
Firstly, we need to press the reset button on the thermal overload relay next to the power distribution
Secondly, we need to check if the distance between the left, right, front, and back of the device is about 60cm, and if the ambient temperature is too high
Finally, we need to clean our condenser with a brush or blow it off
In short, the DI5 alarm indicates compressor overload protection and poor heat dissipation