Step One
Open the right panel of the device and check if there is water in the humidification bucket
Step two
If there is no water in the humidification tank, check if the valve is not closed or if the water pipe is blocked. It is necessary to clean the water pipe, cup, and whether the pumping motor is running normally
Step three
After adding water to the humidification bucket, do not immediately set the humidity level. Instead, set the temperature to high and wait for the water to pass through the humidification tube before setting the humidity level. This is beneficial for preventing the humidification tube from drying out and burning
summary
DI3 alarm is a water shortage alarm, and the humidification pipe is dry burning. Therefore, it is recommended that the equipment needs regular maintenance and treatment, so that the service life of the equipment will become better and better