How to handle the DI3 alarm in the constant temperature and humidity test chamber?
Date: 2024-12-12Read: 11

Step One

Open the right panel of the device and check if there is water in the humidification bucket

Step two

If there is no water in the humidification tank, check if the valve is not closed or if the water pipe is blocked. It is necessary to clean the water pipe, cup, and whether the pumping motor is running normally

Step three

After adding water to the humidification bucket, do not immediately set the humidity level. Instead, set the temperature to high and wait for the water to pass through the humidification tube before setting the humidity level. This is beneficial for preventing the humidification tube from drying out and burning

summary

DI3 alarm is a water shortage alarm, and the humidification pipe is dry burning. Therefore, it is recommended that the equipment needs regular maintenance and treatment, so that the service life of the equipment will become better and better