Functional characteristics of the titration part of the dual tube fully automatic fluorine and chlorine analyzer
Date: 2025-11-19Read: 8

HFFL-6S双管全自动氟氯测定仪.jpg


HFFL-6SDouble tube fully automatic fluorine and chlorine analyzerFunctional characteristics of titration section:

1Equipped with fully automatic titration software, it can be used forWIN10Computer operating system.

2The titration solution can be set, and the titration speed can be set according to the characteristics of the sample.

3It can directly read the titration milliliters and millivolts, and display the titration curve of the titration process; The measurement results are automatically generated.

4The software has a built-in database, and experimental data can be exported, printed, and queried.

5Adopting a high-precision titration system, titration control is more precise.

6Intelligent automatic titration, which can change the titration speed according to the speed of potential changes without manual intervention.

7It has an automatic stop titration function, which automatically stops when the instrument detects the endpoint of the potential, saving resources and time;

8It has a pause function for the experimental process, and one click pause is convenient and practical.

9Equipped with high-speed cleaning function, saving time and effort after the experiment ends.

10It has a stirring adjustment function, and the stirring speed can be adjusted freely.

11Calibration of electrode parameters: It is carried out during the temperature rise sample testing process of the instrument, and is not included in the total testing time. Calibration does not need to be carried out every day. Generally, calibration is only required after long-term use, electrode long-term use, or replacement with a new electrode.

12Analysis time of test solution: Fluorine3-5min; Chlorine about3-5min


In order to thank new and old users for their support and love of our company's products, our team will provide technical support and after-sales service. We will also continuously share some coal knowledge, operating procedures, instrument maintenance, common problems, and precautions with everyone.

If you want to know more aboutCoal quality analysis instrumentFor more details, please closenote Hebi Hengfa Instrument Co., Ltd / ,: