What are the standard regulations for neck strength testing of dental needles
Date: 2025-10-21Read: 13


The dental needle neck strength tester is mainly used to evaluate the neck bending strength, torsional performance, and fatigue life of dental rotating instruments (such as needles, drills), and is in compliance withYY0302-1998TheYY/T0874-2013Waiting for standards.

800580.jpg


Testing standards and instruments

Dental Needle Neck StrengthTesting should followYY 0302.1-2010The Needle Dental Rotating Instruments1part:Steel and Hard Alloy Needle ",YY 0302.2-2016The methods and requirements for testing the neck strength of dental needles in the standards.YY 0302.1-2010The requirements for neck strength in China are as follows:

The needle must not break,Or the deformation should not exceed:

---Steel car needle:0.08mm;

--Hard alloy car needle:0.05mm.

The test should be conducted after the corrosion testYY 0302.1-2010In the standard6.4Requirement to proceed.

High speed rotating dental needle (with a rotational speed of up to20,000-400,000 RPM)Enduring complex stresses in clinical use,Dental Needle Neck StrengthexperimentA professional that meets the standard requirements should be usedDental Needle Neck StrengthtesterDuring testing, it is necessary to ensure that the steel needle is loaded with a load5Deformation after seconds ≤0.08mmTo provide standard basis for testing data....