● Places where direct sunlight shines and near heating appliances
Please choose a place with minimal temperature changes and close to room temperature (23 ℃) as much as possible. If the instrument is installed in a place directly exposed to sunlight or near a heating device, it will have a negative impact on the interior of the instrument.
Places with high levels of oil fumes, steam, moisture, dust, corrosive gases, etc
Oil fumes, steam, moisture, dust, corrosive gases, etc. can have adverse effects on instruments.
● Near the electromagnetic source
Please do not place magnetic instruments or magnets near this instrument. If this instrument is installed near a strong electromagnetic field source, display errors may occur due to the influence of the magnetic field.
To avoid abnormal operation of the instrument, it is necessary to maintain a distance of at least 20cm from the instrument when using the RF generator on site.
● Places that are not convenient for viewing the image
The display section of this instrument uses a 7-inch TFT true color LCD, which may be difficult to see from an extremely biased angle. Therefore, please try to install it in a place where the observer can view it from the front