Classification and Selection Guide for Pressure Calibration Stations
Date: 2025-06-10Read: 12
Pressure calibration tablePressure Calibration Bench is an integrated, high-precision pressure calibration workstation used for batch calibration of pressure instruments (pressure transmitters, pressure gauges, pressure switches, etc.) in laboratories or workshops. Its core features are multi module collaboration, high degree of automation, and high calibration efficiency, suitable for metrology institutions, instrument manufacturing plants, and large industrial enterprises.

1. Classified by pressure medium
type pressure range Applicable instruments feature
Air pressure calibration station -0.1~10 MPa Barometer, pneumatic transmitter Clean and oil-free, suitable for oil free occasions
Hydraulic calibration table 0.1~250 MPa High pressure gauge, hydraulic sensor Stable force value, suitable for ultra-high voltage calibration
Vacuum calibration table -0.1~0 MPa Vacuum gauge, composite pressure gauge Integrated vacuum pump, pumping speed ≥ 1L/min

2. Classify by accuracy level

  • Measurement level: 0.01 level (laboratory standard), high cost

  • Industrial grade: 0.05 grade (for workshop use), with excellent cost-effectiveness

3. Key parameters for selection

parameter reference value Instructions
pressure range -0.1~60MPa (conventional) Covering 120% of the range of the tested meter
stability ±0.005%FS/10min Affects the repeatability of calibration
number of channels 4~8通道 Determine the number of instruments to be calibrated simultaneously
communication protocol HART/RS485/Ethernet Support reading and writing of intelligent instrument parameters
Temperature control function optional Temperature sensitive instruments require environmental temperature control (± 1 ℃)