Product Introduction:
Electrolytic water vapor permeability testerWVTR-AE3 is a professional, efficient, and intelligent testing system suitable for measuring the water vapor permeability of high barrier materials such as plastic films, aluminum foil plastic films, waterproof materials, and metal foils. Expandable test bottles, bags, and other containers.
Product features:
Embedded high-speed microcomputer chip control, simple and efficient human-computer interaction interface, providing users with a comfortable and smooth operating experience;
The design concept of standardization, modularization, and serialization can maximize the satisfaction of users' personalized needs;
7-inch high-definition color LCD screen, real-time display of test data;
● Business testing software, real-time display of test data and curves;
Imported high-speed and high-precision acquisition chips effectively ensure testing accuracy and real-time performance;
One click operation, simple and convenient, learn and master immediately;
● Fully automatic temperature control, capable of arbitrary heating and cooling, with high control accuracy, stability and reliability, and the equipment is not affected by external environmental temperature interference;
Efficient testing with low nitrogen consumption;
● Manual flow control: Testing flow, including carrier gas and moisture, must be manually controlled;
● Manual humidity control: The humidity is manually controlled using mechanical knobs and requires manual intervention;
● Semi automatic testing: Set the testing parameters, manually adjust the flow rate and humidity, and conduct the test automatically. Automatically determine the end of the test and save the test results;
Electrolytic moisture analysis sensor with high measurement accuracy and stable performance;
Precision pipelines and valves, with advanced pipeline flushing technology, ensure the cleanliness of the carrier gas and improve testing efficiency;
Unique sample anti leakage sealing installation technology enhances sample sealing;
Main parameters:
indicator
|
parameter |
| Measurement range (membrane) |
0.01~40 g/(m ² · day) (standard)
0.1-1000 g/(m ² · day) (optional)
|
| Sample size |
3 |
| resolution |
0.001 g/(m²·day) |
| sample size |
108mm×108mm |
| Test area |
50cm² |
| sample thickness |
≤3mm |
| Temperature control range |
15 ℃~55 ℃ (resolution ± 0.01 ℃) |
| temperature control accuracy |
±0.1℃ |
| Moisture control range |
0%RH、35%RH~90%RH、100%RH |
| Carrier gas |
99.999% high-purity nitrogen (gas source to be provided by the user) |
| carrier gas flow rate |
0~200ml/min (manual control) |
| air supply pressure |
≥0.28MPa/40.6psi |
| interface dimensions |
1/8' |
| Correction method |
Standard film calibration |
| host size |
350mm (L)×695 mm (W)×410mm (H) |
| host weight |
60Kg |
| power supply |
AC 220V 50Hz |
Experimental principle:
Using the principle of electrolytic sensor testing, nitrogen with stable relative humidity flows on one side of the film, while dry nitrogen flows on the other side of the film; Due to the presence of humidity gradient, water vapor will diffuse through the film from the high humidity side to the low humidity side; On the low humidity side, the water vapor passing through is carried to the sensor by the flowing dry nitrogen gas. When entering the sensor, an electrical signal of the same proportion is generated. By analyzing and calculating the electrical signal of the sensor, parameters such as the water vapor transmission rate of the sample can be obtained. For packaging containers, dry nitrogen gas flows inside the container while the outside of the container is in a high humidity state.
Compliant with standards:
YBB 00092003、GBT 26253、ASTM F1249、ISO 15106-2、 TAPPI T557、 JIS K7129、ISO 15106-3、GB/T 21529、DIN 53122-2