Operating Manual for DOG-2082YS Fluorescence Dissolved Oxygen Analyzer
Product Specifications
Specifications |
Details |
Product Name |
Industrial Onlineopticsdissolved oxygencontroller |
Shell material |
ABSplastic |
power supply |
90 –260V AC 50/60Hz |
output method |
Two 4-20mA outputs (dissolved oxygen, temperature) |
relay |
5A/250V AC 5A/30V DC |
size |
144×144×104mm |
weight |
0.9kg |
contact method |
Modbus RTU |
measurement range |
0~20.00 mg/L 0~200.00 % -10.0~100.0℃ |
measurement accuracy |
±1%FS ±0.5℃ |
waterproof rating |
IP65 |
The main panel of the fluorescence dissolved oxygen measuring instrument consists of two modules, an LED LCD display module and a button module.
Users can set and adjust the parameters of the instrument through the 5 buttons on the panel.
After the startup animation ends, enter the measurement main interface.
3、 Setting up
Press the 'Settings/Exit' button to enter the password input interface.
communication protocol
Communication parameters:
Baud rate: 4800、 nine600、 one9200(Default is 9600)
Serial data format: 8N1
Equipment address: Fluorescence dissolved oxygen controller defaults to 6
Register definition:
register address |
Register Definition |
R/W |
Instructions |
0 |
temperature |
R |
0.1℃,sint16 |
1 |
dissolved oxygen |
R |
0.01mg/L,uint16 |
2 |
nA |
R |
0.01nA,uint16 |
3 |
saturation |
R |
0.1%,uint16 |
6 |
Instrument type |
R |
The type of dissolved oxygen controller for fluorescence method is 7 |
7 |
software version |
R/W |
Display the current software version, such as 101Then it is ver.1.01 |
8 |
RTUaddress |
R/W |
ModbusMailing address, default fluorescence method dissolved oxygen is 6 |
9 |
Baud rate |
R/W |
4800, nine600, one9200, default is 9600 |
2、 Industrial online fluorescence method dissolved oxygen controller matching electrode parameter table
power supply |
12VDC |
measurement range |
0-20.00mg/L(0-20.00ppm) |
temperature range |
0-45.0℃ |
Measurement progress |
Measurement value ± 3% Temperature ±0.5℃ |
pressure resistance |
0.3MPa |
waterproof rating |