EA2-TX-100 online dew point sensor
Date: 2025-07-25Read: 9

Online Dew Point Sensor Product Introduction:

The design concept of Easidew transmitter is simplicity and ease of use, and its integrated features are designed to simplify installation and usage services as much as possible. The transmitter that is set up and calibrated makes dew point measurement as easy as temperature and pressure measurement, and can be directly connected to your air or gas management and control system.

·Two wire power supply circuit connection

·Dew point unit or ppm moisture content unit

·IP66(NEMA 4)

·Excellent sensor protection

·Range -100~+20 ° C

·Working temperature -40~60 ° C

·Fast response speed

Technical parameters of online dew point sensor:

performance

Dew point range

-Dew point at 100~+20 ° C

Dew point accuracy

± 2 ° C dew point

response time

T95 is 5 minutes (from dry to wet)

repetitiveness

Dew point at 0.5 ° C

input

output signal

4-20 mA (two-wire) active output, customizable dew point across the full range

-100 ~ + 20°C
-148 ~ +68°F
0-3000 ppmV ppmV output or non-standard output range must be specified when placing the order

power supply voltage

12–28 V DC

resistive load

Big 250 Ω @ 12V
500 ohms @ 24 V

Current consumption

20mA large

The impact of power fluctuations

±0.005% RH/V

operating condition requirements

Operating Temperature

-40 ~ +60°C

Operating Humidity

0– RH

work pressure

大 45 MPa (450 barg)

traffic

1 ~ 5 Nl/min, Installed inside the standard sampling sleeve;
0 to 10m/sec, Direct insertion type

temperature coefficient

Temperature compensation has been made within the working temperature range

mechanical properties

Sealing protection level

The corresponding standard for IP66 is BS EN 60529:1992,
And the corresponding standard NEMA 250-2003 for NEMA 4

Shell material

stainless steel

size

L=132mm x ø27mm

Mounting Thread

5/8' - 18 UNF

weight

150g

filter

HDPE protective cap
80 μ m stainless steel protective cap (optional)

electrical connection

See table

interchangeability

Sensors are interchangeable

Fault Overview
(Factory settings)

situation

output

Sensor malfunction

23 mA

Dew point below range

4 mA

Dew point exceeds the range

20 mA