Welcome Customer !

Membership

Help

Shanghai Fatai Precision Instrument Co., Ltd
Custom manufacturer

Main Products:

ybzhan>Article

Shanghai Fatai Precision Instrument Co., Ltd

  • E-mail

    2259938953@qq.com

  • Phone

    18656860316

  • Address

    Room 608, Building 1, No. 348 Huajiang Road, Jiading District

Contact Now
Electrolytic micro water analysis detector
Date: 2025-07-28Read: 9

FT-F31Electrolytic micro water analysis detector, is a trace water analysis instrument developed by our company specifically for detecting the moisture content of highly corrosive gases such as liquid chlorine and hydrogen chloride.

adoptP2O5 electrolysis method, in accordance with national standards such as GBT 5138-2021 Industrial Liquid Chlorine, HGT3941-2007 Determination of Moisture Content in Industrial Liquid Chlorine by Electricity Method, etc

applicationField:

ØChlor alkali industry, medical oxygen, electrolytic aluminum and other industries

ØMeasurement of Trace Water in Various Corrosive Gases

ØMeasurement of Trace Water in Semiconductors and High Purity Gases

ØMeasurement of Trace Water in Special GasesInstrument features:

ØSplit type, the gas path unit and detection unit are designed separately to prevent the sample gas from corroding the sensor

ØThe materials used for components such as the air circuit of the entire machine are corrosion-resistant

ØLarge screen dot matrix formatLCDDisplay, display units are optionalppm、%

ØAutomatic real-time recording of measurement data

Following standards:

ØGB/T 5832.1-2016Gas Analysis Determination of Trace Moisture 第Part 1: Electrolytic Method

ØGBT 5138-2021 Industrial Liquid Chlorine

ØHGT 3941-2007 Determination of moisture content in industrial liquid chlorine - Electricity method

ØGB/T 13277.3-2015Compressed Air 第Part 3: Humidity Measurement Methods

Technical Specifications:

ØMeasurement principle: electrolysis method

ØMeasuring range:01000Ppm Other ranges can be customized

ØDisplay unit:ppm、%

Ødivide distinguish Rate:0.1ppm

Øheavy again Sex: Soil1.5%FS

Østable fixed Sex: Soil1.5%FS

ØResponse time:≤60S

ØWorking environment: Temperature-25°C-55 ° C; humidity ≤ 90% RH

ØSample gas pressure:0.05 MPa ≤ inlet pressure ≤ 0.1 MPa

Øheavy Quantity: Approximately5kg

Øelectricity Source:24V

ØInstrument size:

lDisplay unit260Mm (width) × 160Mm (height) × 300mm (depth)

lAnalysis Unit300Mm (width) ×400Mm (height) ×200mm (depth)

ØInstrument weight:

lDisplay unit approximately1.5kg

lTesting unit approximately8kg