E-mail
2197115204@qq.com
-
Phone
13761613451
-
Address
No. 41, Lane 2688, Yindu Road, Minhang District, Shanghai
Shanghai Heyi Instrument Co., Ltd
2197115204@qq.com
13761613451
No. 41, Lane 2688, Yindu Road, Minhang District, Shanghai
(5) Personal dose alarm device.
1. Purpose: Used for occupational health supervision personnel to issue alarms for doses and dose rates exceeding the set limits.
2. Requirements:
(1) Compliant with the requirements of "Radiation Protection Instruments for Measuring X, γ, Neutron, and § Radiation Personal Dose Equivalents Hp (10) and Hp (0.07) Direct Reading Personal Dose Equivalents" (GB/T 13161-2015);
(2) Measurement range. Cumulative dose equivalent: O.OOpSv~99.99 mSv; Dose equivalent rate: 0.01 μ Sv/h to 10mSv/h;
(3) Alarm function: When the cumulative dose or dose rate exceeds the set threshold, there will be sound, light, and vibration alarms;
(4) Operating environment: -30 ° C to+50C.
3. Precautions for use: Reasonably set the alarm threshold.
(6) Radiation monitoring instrument
1. Purpose: Used for occupational health supervision personnel to detect X and Y radiation dose levels in radiation workplaces.
2. Requirements:
(1) General requirements: The instrument should be easy to operate, portable and handheld;
(2) Measurement range: The low measurable reading of the instrument should not exceed 0.1 μ Sv/h;
(3) The instrument should be able to adapt to pulse radiation field measurement (GBZ201.5);
(4) Operating environment: -10 ° C to+40 ° C;
(5) Can meet the requirements of relevant standards.
3. Precautions for use:
(1) Try to choose devices with shorter response times;
(2) Perform time correction on the detection results based on response time.