Solution for Blood Purification Device Extracorporeal Circulation Blood Circuit Comprehensive Tester
Date: 2025-12-11Read: 12

Blood Purification Device Extracorporeal Circulation Blood Circuit Comprehensive Tester

The comprehensive testing instrument for extracorporeal circulation blood circuit of blood purification device integrates multiple functions, including safety testing of filters and pipeline joints, as well as evaluation of the sealing and flow control of extracorporeal circulation blood circuit. ‌‌

equipmentThe data acquisition system and high-precision sensors used monitor real-time indicators such as pressure and flow during the testing process, and generate detailed performance reports to support data traceability and evaluation. This method reduces the reliance on manual visual inspection, making detection more accurate and data-driven. ‌‌

·one、core functionality

·Pump tube performance testMeasure the flow deviation of the pump tube within the predetermined pressure range to ensure its stability.

·Tightness testingBy conducting positive and negative pressure tests, the sealing of each connection in the blood circuit is checked to prevent leakage.

Blood pathway compliancetest:Quantifying the compliance of blood dialysis by measuring the diameter change of the blood pathway under pressure changes is crucial for ensuring its safety and effectiveness.

2、 Key technical parameters and performance requirements

control system

PLC

operating interface

7Inch color touch screen, language switching between Chinese and English

test project

Pump tube performance; Blood pathway compliance Positive and negative pressure test for tightness

holding time

1s~9999sCan be set

‌data storage‌

≥50,000article

flow sensor

0~400ml/minAccuracy: ±2.5%

‌pressure sensor‌

-100~100kpa,(- 760 mmHg ~ +760 mmHg)‌

Test time

99s/99min(Can be set)

power supply

220V,50hz

threeKey points of maintenance and compliance

‌periodic calibration‌Verify the pressure sensor to ensure data reliability‌;

‌Standard compliance‌:YY 0267-2016

‌safety protection‌Built in overload protection and power-off memory function to prevent accidental interruption of testing‌.

four、Configuration List

host1Taiwan

test software1set

Instruction Manual1this

power cord1root

Certificate of Conformity1copy

Warranty Card1copy

brochure1this