E-mail
mx1718@126.com
-
Phone
18923848017
-
Address
102, Building 206, Gonghe Xinxu, Longhua District, Shenzhen
Shenzhen Meixin Instrument Electronics Co., Ltd
mx1718@126.com
18923848017
102, Building 206, Gonghe Xinxu, Longhua District, Shenzhen
Differential pressure sensor |
resolution |
1Pa |
Sensor range |
±3000Pa |
|
display precision |
±0.5% FS |
|
Test pressure sensor |
display precision |
±1% FS |
display unit |
Test pressure |
KPa |
leakage |
Pa、mL/min、mL/h |
|
pressure regulating valve |
form |
manual |
Leakage display |
5 people |
|
Leakage limit value (standard) |
Below ± 999Pa |
|
input power |
100~240V AC,50/60Hz |
|
Test pressure input |
Use clean air above the test pressure |
|
Drive pressure input |
above0.5MPa clean air |
|
display screen |
7-inch touch screen |
|
Environmental temperature for use |
5~45℃ |
|
humidity |
Below 70% RH |
|
weight |
about10kg (standard specification) |
|
controlI |
Input signals: start, stop, channel |
|
Output signal: under test、 + TC | ||
communication interface |
RS232 or RS485 |
|
USB port |
Record test data downloads and software version updates |
|
network interface |
Remote fault diagnosis, software version updates |
|
Barcode scanner connection |
USB port |
|
An instrument that uses a differential pressure sensor as a pressure sensing element to detect the pressure difference between the measured object and the reference object, in order to determine whether there is leakage and the magnitude of the leakage rate of the measured object
Compared to the tested objectA reference object with the same shape, material, and volume, and whose leakage rate can be ignored compared to the measured object
Under certain pressureThe amount of substance leaked from the inner cavity of the tested object to the outside per unit time
The unit of measurement for pressure isPa,The unit of measurement for leakage rate ismL/min
LC1000 series leak detectorIt is a high-performance instrument that uses differential pressure sensors for leak testing
lUsing touch screen for human-machine information interaction
lDuring the leakage testing process using flow units, the test value of the differential pressure sensor is simultaneously output, which can read the pressure changes inside the testing system at different time periods while reading the leakage amount
lCan provide fromDifferent models of instruments within the testing pressure range of 0.6kPa to 1.8MPa
lCan be connected to a barcode scanner to merge scanning information with test data for storage or output
lThe internal valve body of the instrument adopts an air controlled valve structure, which increases the ventilation volume compared to previous models. The valve body does not generate heat, which improves the accuracy of measurement values while accelerating inflation efficiency
lProvide pressure calibration interface, customers can calibrate on-site