TRD320Integrated (split) electromagnetic flowmeterIt consists of two parts: sensors and converters. It is an inductive instrument used to measure the volumetric flow rate of conductive liquids with a conductivity greater than 5 μ S/cm, based on Faraday's law of electromagnetic induction. In addition to measuring the volumetric flow rate of general conductive liquids, it can also be used to measure the volumetric flow rate of strong corrosive liquids such as strong acids and alkalis, as well as homogeneous liquid-solid two-phase suspensions such as mud, slurry, and pulp.
l Product Features:
1. Spraying zinc on the surface of the pipe can also provide protection in harsh environments, and its anti-corrosion performance is superior;
2. The sensor housing is made of precision aluminum mold, which fits tightly, has no welding as a whole, is easy to assemble and disassemble, and is convenient for maintenance;
3. High definition backlit LCD display, easy to use and operate. Can display forward, reverse, and differential cumulative calculation quantities;
4. Thickened lining, curved electrode, * with good corrosion resistance and wear resistance, stronger adaptability to mixed media;
5. The header can rotate 360 degrees, making it more convenient to display in situations with limited installation space;
6. There are no obstruction components or pressure loss inside the measuring tube, and the straight pipe section requires low requirements. Has unique adaptability to slurry measurement;
7. Adopting SMD devices and surface mount technology, the circuit reliability is high; The magnetic field distribution of the sensor is uniform and the signal is strong.
8.
On site, the range can be modified online according to the actual needs of the user to better match with the collection terminal;
9. A 16 bit embedded microprocessor improves the stability of flow measurement and has low power consumption;
10. Equipped with digital communication signal outputs such as RS485, RS232, Hart, and Modbus Profibus DP (optional)
11. The measurement results are independent of physical parameters such as fluid pressure, temperature, density, viscosity, etc; Equipped with self checking and alarm functions;
12. Hour total recording function (optional); There is an internal power-off clock that can record 16 power-off times (optional);
13. Infrared handheld operator, 115KHZ communication speed, all functions of long-distance non-contact operation converter (optional);
Ultra low EMI switching power supply, suitable for a wide range of power supply voltage changes, with good EMI resistance performance. Overall stable performance and stronger anti-interference ability;
Note: Due to differences in instrument configuration and functionality, the reference price may deviate significantly from the actual price. For selection pricing and delivery time, please consult our sales consultant by phone. Thank you!