Instructions for using the infrared temperature screening device
Date: 2020-02-21Read: 10

AOD-HT1403AS(Upgraded) Infrared Body Temperature Screening Device User Manual

Thank you for purchasing and using the product produced by AioudeAOD-HT1403A SInfrared temperature screening device. In order for you to better use our product, please read this user manual carefully before use.

eye record

1.Product Introduction 2

2.Basic working principle 2

3.Product features and scope of application 2

3.1Product Features 2

3.2Product scope of application 2

4.Main technical parameters of the product 2

5.Basic composition and external structure (see figure) 3

6.Auxiliary optional function 3

7. Equipment installation 3

7.1 Install the host and internal circuit board 4

7.2 Install sensors and other components 4

7.3 bracket installation 5

8.software installation 6

8.1installUSB-485The Drive 6

8.2Search for serial port. 7

9.Job Description 7

9.1power on 7

9.2 .Setting of thermometer 8

9.3 measurement 8

10. WIFIRouter (optional) 9

11. Precautions for use 9

12.Packing List 9

13.Maintenance and Storage 9

14Warranty Instructions 10

1. Product Introduction

AOD-HT1403A SInfrared temperature screening device (hereinafter referred to as screening device) is a measuring instrument that uses the principle of infrared reception to measure human body temperature. When in use, when the human body passes through the screening device, a glance at the lens can quickly and accurately measure the body temperature.

2.Basic working principle

All objects with temperatures above zero will emit a certain proportion of infrared radiation energy based on their own temperature. The magnitude of radiation energy and its distribution according to wavelength are closely related to its surface temperature. The human body temperature is(36℃-37℃)The infrared wavelength of radiation is9~13um.Based on this principle, accurate measurement of the surface temperature of the human forehead and correction of the temperature difference between the forehead and the actual body temperature can be used to display the accurate body temperature.

3.Product features and scope of application

3.1Product Features

1. Multi point infrared rapid scanning to cope with different heights of pedestrian flow, suitable for temperature scanning of people in the channel;

2. adopt8-14uThe infrared range of the human body allows for longer measurement distances;

3. Voice temperature broadcast, high temperature voice alarm function;

4. Provide automatic scanning to detect abnormal body temperature mode and trigger measurement of nearby body temperature mode;

5. useDCPower interface, safer to use.

3.2Product scope of application

AOD-HT1403A SScreening device Widely used in schools, factories, stations, airports, customs, cinemas, conference halls, supermarkets and other places that require rapid scanning of human body temperature.

4.Main technical parameters of the product

Normal working conditions

ambient temperature.:10—35℃

relative humidity:≤85%

power supply:DC 9V≥500mAadapter

basic size

(1500-2200)mm*360mm*520mm

gross weight

15Kgs

resolution

0.1℃

measurement range

34℃~43℃

measurement error

±0.4℃

measurement repeatability

±0.2℃

measure time

≤1 S

Measurement distance:

20~40CM

Software operating environment

WINXP(higherPCVersion compatibility requires customers to develop their own client)

5.Basic composition and external structure (see figure)

Back interface board

6.Auxiliary optional function

1. High brightness large digital tubeLEDExternal display, with a long visible distance;

2. provideWIFITransmissionusbInterface,1A computer can connect to both single and multiple devices8The thermometer can also be used separately;

7. Equipment installation

The complete shipment of the host can be ignored7.1-7.2

7.1 Install the host and internal circuit board

Please arrange the host and internal circuit board in the following order.

7.1.1After removing the host, remove the top cover of the host, and then pull out the bottom plate to see the motherboard and internal wiring. According to the diagram1Connect the internal circuit to the motherboard at the indicated position and check itWIFIAre the modules and antennas securely plugged in.

image1 Motherboard wiring

7.2 Install sensors and other components

Note that the sensor is a sensitive component, and be careful not to drop or disassemble the sensor itself during installation. The probe must be installed on a vertical panel and fixed with screws, in one set6One.

7.2.1Within the same group10Each probe is set to1-6Number, setting method reference diagram2.

image2 Back view of probe

7.2.2check6Does the setting of the probe comply with the diagram2Fix the probe onto the host with screws, and then use6Connect the probes one by one and connect them to the motherboard through the wiring. Probe press1-6Place from top to bottom for easy inspection. Attention: When fixing the probe, do not fix it too firmly with screws to avoid damaging the probe. reference diagram3.

image3 Probe fixation and wiring

7.2.3Power on inspection of the entire machine.

a)Input at the bottom of the host9Vload500mAThe above DC power supply, the host starts working, the host'sUSBconnect toPCofUSBInterface, or useWIFI,PCofUSBAttached with the mouth connectionWIFIRouter and run communication software, switch the serial port of the communication software to the correct serial port(USBRefer to section for the installation of the driver2Section).

b)Automatic inspection of equipment after power on6A probe, hostLEDIf the digital display shows the probe number and the inspection is passed, it will be displayedGoodIf it does not pass, the main control board will display the probe number and then showEx-z(xFor the probe number,zFor fault codes.

Z=1:Unable to find the probe board, other than probe sensor malfunction.

Z=1Please check if there are any duplicate or connection issues with the coding settings of the probe board.

c)The communication software should be able to receive the measurement process and data. If it cannot receive, please check the connection and serial port settings(WIFINeed to be powered on1It takes more than minutes to connect online.

7.3 bracket installation(As shown in the figure)

1. First, install the sliding wheel of the base.

2. Translate into English2The root pillar is fixed to the base with screws.

3Thread the host from top to bottom onto the column and secure it to the column with screws at an appropriate height (screws on the back).

image4 bracket installation

8.software installation

If not selectedWIFIYou don't need to read this section.

8.1installUSB-485The Drive

useUSBWire connection equipment andPCYes, it is,PCWill prompt to install oneUSB-SERIALDriver software, the driver software is for deliveryCH341SER. EXEClick on the software directly to install it. runCH341SER. EXERear display image5Screen, clickInstallInstall the driver. After installation, insert the next timeUSB-485toPCThe system will automatically install drivers afterwards.

image5 usb-485Driver installation

8.2Search for serial port.

installUSB-485After driving, a virtual serial port is generated, and the serial port number needs to be enteredSerialOnly then can it work normally. Refer to the following for serial port search.

Click on 'My Computer'(My Computer)/hardware(Hardware) /Device Manager(Device Manager)/Port(Port)(COMandLPT)”, View“USB-SERIAL CH340(COM?)”of That is the serial port number. If you need to change the serial port number, double-click the line and a property dialog box will pop up, select "Port Settings/Advanced(A)…”Go to the port selection page and select“COMPort number(P)”Select the desired port from the dropdown menu, usually chooseCOM4Then press confirm, ignore other warnings and raise the issue. Refer to the figure6

image6 USBSerial port search

9.Job Description

9.1power on

Automatic inspection of equipment after power on6A probe, hostLEDIf the digital display shows the probe number and the inspection is passed, it will be displayedGoodIf it does not pass, the main control board will display the probe number and then showEx-z(xFor the probe number,zFor fault codes. Display panel as shown in the figure7

image7 display panel

Definition of indicator light:ALRRed, lights up when the measured body temperature exceeds the set alarm temperature

COMMGreen, flashing when there is data communication

WIFIWhen connected toWIFIThe router lights up.

9.2 . Screening deviceThe settings

For secondary use of the screening device, it may be necessary to set it according to the usage scenario and mode. By pressing the buttons, the screening device can set alarm temperature, voice, and measurement modes.

9.2.1Set alarm temperature

settle downSETkey2More than a second, enterF—1After releasing the button, the setting value will be displayedUPandDOWNThe key can change the alarm value, and press the button to change itENTERSave key.

When the measurement result exceeds the temperature, the voice broadcast will continue after the temperature is reported“Temperature is higher/High body temperature.

9.2.2Set up voice alarm

enterF—1Press again laterSETKey, enterF--2After releasing the button, the voice setting value will be displayedUPandDOWNThe key can change the voice, and press it to change the soundENTERSave key.

Voice value0: No voice

Voice value1: English pronunciation

Voice value2Chinese pronunciation

9.2.3Set voice volume

enterF—1Press again later2timeSETKey, enterF--3After releasing the button, the voice volume value will be displayedUPandDOWNThe key can change the volume,10For high volume,0To silence.

9.2.4Set measurement mode

enterF—1Press again later3timeSETKey, enterF--4After releasing the button, the measurement mode will be displayedUPandDOWNThe key can change the measurement mode:

pattern0Automatically scan temperature, regardless of whether someone has passed by or not. If a high temperature is detected during the scanning process, the temperature will be displayed and transmitted toPCIf it exceeds the alarm temperature, an alarm will be triggered;

pattern1Weight Scale Trigger Mode(Need to purchase a weight scale);

pattern2Infrared sensing trigger mode, with an internal infrared sensor that only measures when someone is detected, and the measurement results are always uploaded toPCIf the temperature exceeds the alarm temperature, an alarm will be triggered.

9.3 measurement

Can be used during measurementWIFIConnect orUSBDirect connection of wiresUSBMouth, runningPCAfter the software is installed, select the correct serial port for monitoring and collecting measurement data.

When the subject passes by the thermometer, standing slightly in front of the thermometer can quickly measure the subject's temperature. If the thermometer is working in mode2The thermometer will upload the measurement resultsPCAnd alarm for body temperature exceeding the alarm temperature. If working in mode0Only information exceeding the alarm temperature will be uploadedPC.

9.4 Measurement issues and solutions.

9.4.1The thermometer is not working: check if the power is plugged in and if the power switch is turned on,12Is the connecting cable installed.

9.4.2Always unable to measure: Check if the measurement mode is set correctly, if it is a mode2Check if the infrared human body sensing module is not installed. If it is in mode1It must be equipped with a matching weight scale and measured while standing on the scale.

9.4.3Inaccurate temperature value or inability to measure temperature: The temperature depends on the forehead temperature, and if there is sweat or hair covering the forehead, it may cause inaccurate measurement; The forehead is not aligned with the probe; Too far from the thermometer (do not exceed)40CM).

10. WIFIRouter (optional)

WIFIRouters are used in conjunction with devices to enable them toPCRealize wireless data transmission between them. oneWIFIThe router can connect to large8Equipment, the data of the equipment is sent wirelessly toPCThe data is distinguished from its source by device encoding in the format.

WIFIThe use of a router is quite simple, just useUSBConnect the router andPCJust enough,PCWill prompt to install oneUSB-SERIALThe driver is the same as the one prompted when the device is connected. By setting the driver serial port, it is possible toWIFIThe serial port of the software needs to be changed to match the virtual serial port of the device to avoid different connection methods.

11. Precautions for use

11.1Do not install it in areas with air vents, direct sunlight, or rain, and do not install it opposite the probe(10Do not install in places with high temperature heat sources within meters, such as outdoor air conditioning units, black objects that can be exposed to sunlight, etc., and do not install in places with large temperature changes, such as air conditioning vents or entrances and exits to air conditioning rooms;

11.2The instrument should be placed in the environment during use30The person being measured should stay in a similar environment for more than minutes5For more than minutes, the tested person's forehead should not have any obstructions or sweat, and should not carry high-temperature objects to avoid affecting the measurement results.

11.3The person being measured should be away from the instrument during the measurement20-40cmA slight pause in the area will result in more accurate measurements.

11.4Usage pattern0At that time, the ambient temperature was32Below Celsius degrees.

12.Packing List

host

support

base

screw

wheel

knob

adapter

Large digital tube machine (optional)

Instruction Manual

1Taiwan

2root

1Taiwan

8a

4a

2a

1a

1a

1this

13.Maintenance and Storage

  • The probe protective lens is a vulnerable part of the screening device, so it must be carefully protected.
  • This product must be handled with care and should be avoided from being stored in excessively humid, high temperature, or direct sunlight.
  • Lens cleaning: The probe lens must be carefully protected by gently wiping it with a cotton swab or soft cloth dipped in alcohol.
  • Clean the shell by wiping it clean with a damp sponge or soft cloth.
  • Please keep the power interface dry when using the power supply and do not place it in a place that is easily accessible to children to prevent electric shock.
  • This product should not be soaked in water or directly exposed to sunlight.
  • Do not drop or collide the product, as it may cause damage.

The screening device should be stored at a temperature of-25~55℃, relative humidity not exceeding90%An indoor environment free from corrosive gases and with good ventilation. During transportation and loading and unloading, measures should be taken to prevent throwing, heavy pressure, rain, and trampling, and to stack according to the direction indicated on the packaging box.