E-mail
zzcc8162@zzgwsit.cn
- Phone
-
Address
1119 Science Avenue, Shangjie District, Zhengzhou City
Zhengzhou Changcheng Technology Industry and Trade Co., Ltd
zzcc8162@zzgwsit.cn
1119 Science Avenue, Shangjie District, Zhengzhou City
The entire device consists of a refrigeration system, a circulation system, and a control system.
The refrigeration system consists of a compressor, condenser, evaporator, and capillary tube.
compressor:Piston compressor, small size, light weight, smooth operation, low noise, high reliability, high energy efficiency.
Condenser:adoptAir-cooled condenser,Internally, oxygen free copper threaded pipes are used for heat exchangeefficiencyHigh,Easy to clean and maintain.
Throttle mechanism: using capillary tube, stable and reliable performance.
Evaporator: Adopting304Stainless steel coil evaporator with high heat transfer efficiency; The evaporator is installed in the liquid storage tank, and it will not be damaged when the refrigerant freezes, making it safe and reliable.
Adopting low temperature resistanceStainless steel shielded pump is used to transport refrigerant, and the storage tank and circulation pipeline adopt304stainless steelThere is a safety label indicating the maximum liquid filling amount on the storage tank.
Adopting LCD controller, fully digital Chinese display, temperature display accuracy0.1℃, 操作方便、安全、可靠,控制器符合CESecurity certification requirements.Communication adoptsRS485Interface, followingMODBUS RTUagreement.

2、 Main technical parameters of DLSB-5/20A new low-temperature coolant circulation pump
item eye |
Performance and parameters |
|
Product Model |
DLSB-5/20B |
|
No load temperature(℃) |
-20 |
|
*Temperature range for use(℃) |
-20 ~ 25 |
|
Storage tank capacity(L) |
5 |
|
controller |
LCD (Liquid Crystal Display) |
|
temperature stability(℃) |
±2 |
|
Sensor indexing number |
Pt100 |
|
safety protection |
Delay, overcurrent, overheating, grounding protection |
|
rated power(W) |
635 |
|
|
**manufacture cold quantity (W) |
10℃ |
650 |
0℃ |
580 |
|
-10℃ |
490 |
|
-20℃ |
280 |
|
refrigerant |
R22 |
|
|
follow ring pump |
Power(W) |
100 |
traffic(L/min) |
20 |
|
pressure(bar) |
0.4 |
|
Protection level |
IP20 |
|
pollution level |
2level |
|
ambient temperature(℃) |
5 ~ 35 |
|
Relative humidity of the environment(%) |
≤ 70 |
|
power supply |
220 - 240V~,50Hz |
|
Refrigerant outlet and return port |
1/2″ |
|
connect the hose |
Silicone rubber hose:Φ13×Φ9mm |
|
Material of storage tank |
304stainless steel |
|
Overall dimensions(mm) |
335W×473D×684H |
|
net weight(kg) |
41 |
|
|
* The upper limit of the temperature range should meet the following requirementsTmax≤(Room temperature)-5)℃. ** During testing, the ambient temperature is25℃. | ||