CCZG-2 Explosion proof Individualdust sampler
CCZG-2 Explosion proof Individualdust sampler
1CCZG-2 Explosion proof Individual Dust SamplerOverview
CCZG—2Type individual dust sampler is an instrument used to measure the average concentration of dust in the air within a work shift. This instrument is based onMT162-95General Technical Conditions for Workplace Dust SamplersQ/FY-02-2005Enterprise Standards for Dust Samplers andGB3836.4-2000In the standardExibIThe explosion-proof design of the level (intrinsic safety type) has the characteristics of small sampling airflow pulsation, large load capacity, stable flow rate, simple operation, safety and reliability, and easy portability. The separation efficiency of respiratory dust meets the requirements“BMRC”The international standard curve requirements are particularly suitable for use in coal mines and other workplaces containing explosive gases.
IICCZG-2 Explosion proof Individual Dust SamplerSpecification
1. Sampling flow rate: Respiratory dust sampling:3L/min
2Sampling flow error: ≤2.5%
3Sampling flow stability: ≤5%
4Load capacity: ≥1000Pa
5Continuous working time: ≥10h
6Work noise: ≤65dB (A)
7Accuracy of flowmeter: not less than2.5%
8Sampling time error: 5minWithin no more than1.0 s
9Average flow velocity at the sampling port:0.5~4.0m/s
10Sampling efficiency: compliant“BMRC”Sampling efficiency curve
11External dimensions:120×80×42(m m)
12Weight:450g
13The instrument battery adoptsNi-H 1.3A/h,5Section, series connection6only0.51Ω/3WThe current limiting resistor is sealed with epoxy resin and placed inside the battery box. Its output voltageU0:7VOutput currentI0:2AConstitutes an intrinsic safety battery component.
14Explosion proof type of instrument: mining intrinsic safety type, explosion-proof mark:ExibⅠ。
15、Coal mine shafts containing gas or coal dust explosion hazards