Capacitive pressure gauge

NegotiableUpdate on 12/08
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
Capacitive pressure gauge, suitable for: vacuum furnaces, vacuum packaging, plasma sterilizers, freeze-drying, vacuum distillation, petrochemical industry, semiconductor industry, PECVD/ALD process equipment, plasma etching process equipment.
Product Details

Capacitive pressure gaugeUsing highly stable capacitive sensitive components, it has excellent overvoltage resistance and long-term stability, suitable for precision vacuum measurement that is crucial for process control in the photovoltaic, semiconductor, and industrial markets. Advanced manufacturing technology and processes ensure repeatability in application scenarios. The working voltage of this pressure gauge is 24V DC, with a standard (0~10) V DC analog output signal, full-scale measurement, and a usable range of 1Torr (133Pa) to 10Torr (1.33kPa).

Main Features

Industry standard connector

High precision: standard ± 0.5% reading, optional ± 0.25% reading

Precision membrane ensures excellent performance

Wide range of working temperature compensation

Low circuit noise and fast response

Not easily affected by changes in environmental conditions

Excellent anti overvoltage design

Overload and burst pressure rating

Specifications

range 10Torr. A/1.33kPa. A
precision ± 0.5% reading; ± 0.25% reading (optional)
Working Voltage (18~24)V DC
output signal (0~10)V DC
Operating Temperature (10~50)℃
measuring medium gas
electrical connector 9-pin analog interface
process connection KF16
Power on preheating time 10min
Shell material stainless steel
Liquid receiving material 镍基合金

Wiring Definition

电容式压力计

Capacitive pressure gaugeCalibration operation instructions

电容式压力计

Calibration preparation

1. Remove the window protection buckle corresponding to 0.5T, 5T, and 10T adjustment;

Prepare a small screwdriver with a straight edge.

Calibration operation

1. Adjust the external air pressure to 0.5T, wait for 1 minute for the air pressure to stabilize, and then use a small screwdriver to rotate the 0.5T potentiometer clockwise for half a turn and then back for half a turn;

Adjust the external air pressure to 5T, wait for 1 minute for the air pressure to stabilize, and then use a small screwdriver to turn the 5T button;

Adjust the external air pressure to 10T, wait for 1 minute for the air pressure to stabilize, then use a small screwdriver to rotate the 10T potentiometer clockwise for half a turn and then back for half a turn;

After completing the above three steps, the calibration is completed.

Precautions for calibration operation

During the calibration process, it is necessary to wait for 1 minute for the external air pressure to stabilize before rotating the 0.5T and 10T potentiometers or turning the 5T button. After the 10T calibration is completed, the transmitter automatically saves the recorded data.