Introduction: 1. Main parameters of the product; 1. Sterilizer capacity: 0.3M 600 * 1000mm 2. Sterilization temperature ℃: 105-132 ℃ (1 ℃) 3. Working pressure Mpa: 0.15Mpa~0.20Mpa 4. Power: 12KW/380V, 50Hz 5. Each pot consumes 6-8 degrees of electricity (03 type 126 ℃, sterilization time 30 minutes)
1、 Main parameters of the product;
1. Sterilizer capacity: 0.3M ³ φ 600 * 1000mm
2. Sterilization temperature ℃: 105-132 ℃ (± 1 ℃)
3. Working pressure Mpa: 0.15Mpa~0.20Mpa
4. Power: 12KW/380V, 50Hz
5. Each pot consumes 6-8 degrees of electricity (03 type 126 ℃, sterilization time 30 minutes), with excellent insulation effect (equipment surface temperature<45 ℃, ambient temperature<2 ℃), no wet packaging, and can be dried in 10 minutes.
6. Material: The material of the 60B sterilization chamber is 304 stainless steel.
7. YZM-60 automatic control pressure steam sterilizer performance.
Sterilization temperature:
The maximum sterilization temperature can reach 132 ℃ (0.20Mpa);
2. Direct temperature control and digital display;
3. Stepless temperature regulation (105~132 ℃).
Control mode: Automatic control throughout the entire process (including automatic cooling, automatic constant temperature and pressure during sterilization, timing, completion, and fault alarm prompts).
Security protection:
1. Water shortage and power outage protection;
2. Overpressure power-off protection;
3. Over temperature power-off protection;
4. Overpressure steam release protection;
5. Quick opening door safety interlock device;
6. Cabinet style enclosed appearance design, equipment surface temperature<45 ℃.
2、 Technical characteristics
1. This device applies computer technology and has a built-in sterilization program with variable parameters, achieving automatic control of the entire sterilization process and ensuring sterilization effectiveness.
2. To meet the sterilization requirements of some special items, manual control is set up for sterilization, which is freely controlled by the operator.
3. The sterilization temperature is controlled in real-time and displayed digitally by reading data directly from the sterilization room through sensors. Users can intuitively read temperature and time.
4. The low water level automatic power-off device will give an alarm prompt when the water level is too low.