Dew point transmitter EA2-TX-100The design concept of Easidew transmitter is simplicity and ease of use, and its integrated features are designed to simplify installation and usage services as much as possible. The transmitter that has been set up and calibrated makes dew point measurement as easy as temperature and pressure measurement, and can be directly integrated into your air or gas management and control system.
Product Highlights

| Easidew, Easidew M12, and Easidew 34 transmitters | Easew I.S. transmitter | |
| product | ||
| range | -Dew point from 110 to+20 ℃; -Dew point ranging from 100 to+20 ℃; Non standard range can be provided upon request | |
| precision | ± 2 ° C dew point* | |
| response time | 5 minutes to T95 (dry to wet) | |
| repetitiveness | Dew point at 0.5 ° C | |
| verify | Traceable 13 point verification certificate | |
| Electrical parameters | ||
| output signal | 4-20mA (2-wire connection, current source); User configurable range Easidew M12: RS485 Modbus RTU | |
| output | Dew point or humidity | |
| Analog output proportional range | Dew point: -110 to+20 ° C;; Humidity in gas: 0 to 3000 ppmv. ppmV | |
| power supply voltage | 12 to 28 V DC | |
| load resistor | Max 250 Q @ 14 V (500Ω @ 24 V) | |
| current consumption | 23 mA max, Depends on the output signal | |
| CE conformity UL61010-1 & CAN/CSA C22.2 No. 61010-1 | Licensed | |
| working conditions | ||
| Operating Temperature | -40 - +60°C | |
| Compensation temperature range | -20 - +50°C | |
| Storage temperature | -40 - +60°C | |
| operating pressure | 大 45 MPa (450 barg) | |
| traffic | 1 - 5Nl/min, Installed in a standard sampling sleeve; 0-10m/sec direct insertion | |
| mechanical parameters | ||
| Protection level | IP66 according to BS EN 60529:1992 standard; NEMA 4 protection specified in NEMA 250-2003 standard Easidew M12: IP65 | |
| Intrinsic Safety Zone Certificate | ATEX: II 1 G Ex ia IIC T4 Ga(-20 至 +70 °C) IECEx: EX ia IIC T4 Ga (-20 to +70 °C) TC TR@ 0Ex ia IICT4 Ga (-20 to +70 °C) FM: Class I, Division 1, Groups A B C D, T4 CCSAUS: Class I, Division 1, Groups A B C D, T4 | |
| Oxygen service | Optional: Oxygen enriched cleaning service | |
| housing material | 316 stainless steel | |
| size | MiniDIN 43650 Type C L=132mm x Ø 45mm (with connector cable) Easew M12: M12 5-pin L=155mm x Ø 45mm (with connector cable) | |
| Filter (sensor protection) | Standard configuration: HDPE<10 μ m Optional: 316 stainless steel sintering protection<80 μ m | |
| process connection | Easidew: 5/8” - 18 UNF Easidew 34: 3/4” - 16 UNF Easidew M12: 5/8” -18 UNF, 3/4” - 16 UNF, G1/2” BSP | 5/8” -18 UNF, 3/4” - 16 UNF, G1/2” BSP |
| weight | 150g | |
| electrical connection | Easidew:Easidew: MiniDIN 43650 C 型 Easidew M12: 5 pins (code A) | MiniDin 43650 Type C |
| Electrical connector | Standard connector Easew M12: Optional 0.8, 2, and 5-meter M12 connectors/cables | |
| Diagnostic conditions (factory programming) | condition Sensor malfunction Below the lower limit Exceed the limit output 23 mA 4 mA 20 mA | |
| Permitted isolated safety barrier | KFD2-CR-EX1.20200 KFD2-CR-EX1.30200 KFD0-CS-EX1.50P KFD0-CS-EX2.50P KFD2-STC4-EX1.H MTL5041 MTL5040 | |
| *Overcompensation temperature range | ||