Diaphragm sealed pressure transmitter (convex diaphragm shape)

NegotiableUpdate on 12/02
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
Sealing diaphragm is used to prevent the medium in the pipeline from directly entering the pressure sensor component of the differential pressure transmitter. It is connected to the transmitter by a capillary tube filled with fluid. EJA438W and EJA438N diaphragm sealed pressure transmitters are used to measure the pressure of gases, liquids, and vapors, and then convert them into 4-20mADC current signals for output. Diaphragm sealed pressure transmitter (convex diaphragm shape)
Product Details

Product Name:Diaphragm sealed pressure transmitter (convex diaphragm shape)

Product features: The sealing diaphragm is used to prevent the medium in the pipeline from directly entering the pressure sensor component of the differential pressure transmitter. It is connected to the transmitter by a capillary tube filled with fluid. EJA438W and EJA438N diaphragm sealed pressure transmitters are used to measure the pressure of gases, liquids, and vapors, and then convert them into 4-20mADC current signals for output

Diaphragm sealed pressure transmitter (convex diaphragm shape)Introduction:

Sealing diaphragm is used to prevent the medium in the pipeline from directly entering the pressure sensor component of the differential pressure transmitter. It is connected to the transmitter by a capillary tube filled with fluid. EJA438W and EJA438N diaphragm sealed pressure transmitters are used to measure the pressure of gases, liquids, and vapors, and then convert them into 4-20mADC current signals for output. EJA438W and EJA438N can also communicate with each other through the BRAN controller, CENTUM CS/μ XL, or HART 275 controller for setting and monitoring purposes.

Application

Type

model

Membrane box

Range (MPa)

The biggest work pressure

Pressure (diaphragm sealed)

Convex membrane

EJA438N

A

0.06-3

Based on flange specifications

B

0.46-7

Flat membrane

EJA438W

A

0.06-3

B

0.46-14

List of Selection for Diaphragm Sealed Pressure Transmitters

Basic specification code: EJA438N-DASJ12GA-AB01-92NN

Item 1: Output signal

-D

4-20mA DC/BRAN protocol digital communication

-E

4-20mA DC/HART protocol digital communication

-F

FF fieldbus communication

Item 2: Measurement range (membrane box)

A

0.06-3MPa(0.6-30kgf/cm2)

B

0.46-7MPa(4.6-70kgf/cm2)

Item 3: Material of liquid receiving parts

S

[Diaphragm] JIS SUS316L [Tube] JIS SUS316L [Other] JIS SUS316

Item 4: Flange specifications

J1

JIS 10K

J2

JIS 20K

J4

JIS 40K

A1

ANSI 150

A2

ANSI 300

P1

JPI 150

P2

JPI 300

D2

DIN PN10/16

D4

DIN PN25/40

G2

GB PN10/16

G4

GB PN25/40

G6

GB PN100

Item 5: Diaphragm protrusion length (X2)

2

X2=50mm

4

X2=100mm

6

X2=150mm

Item 6: Flange size/material

G

100mm(4')/JIS S25C

H

100mm(4')/JIS SUS304

J

100mm(4')/JIS SUS316

Item 7: Material of bolts and nuts

A

JIS SCM435

B

JIS SUS630

Item 8: Filling solution

-A

Generally, silicone oil is used for liquid contact at a temperature of -10-250 ℃ and an ambient temperature of -10-60 ℃

-B

Generally, silicone oil is used for liquid contact at a temperature of -30-180 ℃ and an ambient temperature of -15-60 ℃

-C

High temperature (silicone oil) contact temperature 10-300 ℃ ambient temperature 10-60 ℃

-D

Prohibited oil (fluorine oil) contact temperature -20-120 ℃ ambient temperature -10-60 ℃

-E

Low temperature (ethylene glycol) contact temperature -50-100 ℃, ambient temperature -40-60 ℃

Item 9:

B

Always for B

Item 10: Capillary length (m)

01

Capillary length 1m

02

Capillary length 2m

03

Capillary length 3m

04

Capillary length 4m

05

Capillary length 5m

06

Capillary length 6m

07

Capillary length 7m

08

Capillary length 8m

09

Capillary length 9m

10

Capillary length 10m

Item 11: Installation Method

-9

Horizontal piping connection, with the left high-pressure capillary tube leading out from the front

Item 12: Electrical Interface

0

G1/2 internal thread, one electrical interface

2

1/2NPT internal thread, two electrical interfaces, no blind plugs

3

Pg13.5 internal thread, two electrical interfaces, no blind plugs

4

M20 internal thread, two electrical interfaces, no blind plugs

5

G1/2 internal thread, two electrical interfaces, one blind plug

7

1/2NPT internal thread, two electrical interfaces, one blind plug

8

Pg13.5 internal thread, two electrical interfaces, one blind plug

9

M20 internal thread, two electrical interfaces, one blind plug

Item 13: Hidden Display Table

D

Digital display table

E

Digital display table with measurement range setting button

N


Item 14: Installing the bracket

A

JIS SECC 2 "pipe installation (flat bracket shape)

B

JIS SUS304 2 "pipe installation (flat bracket shape)

N


Additional functions of diaphragm sealed pressure transmitter

Item 1: Explosion proof type, intrinsic safety type

no


NF1

NEPSI Explosion proof Permit: d Ⅱ CT6 Environmental Temperature: -40-60 ℃ Electrical Interface: 1/2NPT Internal Thread

JF3

JIS 隔爆 Ex ds Ⅱ C T4X

JF35

JIS 隔爆 Ex ds Ⅱ C T4X

FF1

FM Explosion proof: Class I, II&III, Zone 1&2, Groups B, C, D, E, F&G NEMA 4X T6

FF15

FM Explosion proof: Class I, II&III, Zone 1&2, Groups B, C, D, E, F&G NEMA 4X T6

KF1

CENELEC explosion-proof: EEx d Ⅱ C T4, T5, T6 maximum liquid contact temperature: T4-135 ℃, T5-100 ℃, T6-85 ℃

CF1

CSA Explosion proof: Class I, II&III, Zone 1&2, Groups B, C, D, E, F&G Type 4X T4, T5, T6

CF15

CSA Explosion proof: Class I, II&III, Zone 1&2, Groups B, C, D, E, F&G Type 4X T4, T5, T6

NS1

NEPSI intrinsic safety license: ia Ⅱ CT4 environmental temperature: -40-60 ℃ electrical interface: 1/2NPT internal thread

JS3

JIS 本安: Exia Ⅱ C T4

FS1

FM intrinsic safety: Class I, II&III, Zone 1, Groups A, B, C, D, E, F, G, T4

KS1

CENELEC intrinsic safety: EEx ia Ⅱ C T4

CS1

CSA intrinsic safety: Class I, II&III, Zone 1, Groups A, B, C, D, E, F, G T4

FU1

Including FF1 and FS1

KU1

Including KF1, KS1, and N-type KEMAN EX nAII CT4

CU1

Including CF1 and CS1

SU1

SAA explosion-proof, intrinsic safety, and non sparking types: Ex d Ⅱ T6/T5/T4 IP67, Exia Ⅱ CT4 IP67, Exn Ⅱ CT4 IP67

KF5

CENELEC 隔爆: EEx d Ⅱ C T4、T5、T6

KS5

CENELEC (KEMA) intrinsic safety explosion-proof: : EEx ia Ⅱ CT4; : EEx ia Ⅱ CT4/EEx ia Ⅱ BT4

Item 2: Japanese standard explosion-proof joint

no


G11

Power interface: G1/2 internal thread 1 available cable outer diameter: 8.5-11mm

G12

Power interface: G1/2 internal thread 2 available cable outer diameter: 8.5-11mm

Item 3: Explosion proof sealing joint

no


G71

Wiring port: 1/2NPT 1 steel pipe wiring method Applicable cable outer diameter: Φ 8.5 ± 0.5

G72

Wiring port: 1/2NPT 1 cable wiring method Applicable cable outer diameter: Φ 8.5 ± 0.5

G81

Wiring port: 1/2NPT 2 steel pipes Wiring method Applicable cable outer diameter: Φ 8.5 ± 0.5

G82

Wiring port: 1/2NPT 2 cable wiring method Applicable cable outer diameter: Φ 8.5 ± 0.5

Item 4: Painting and Color Change

no


P1

Only applicable to transmitter housing

P2

Only applicable to transmitter housing

P7

Only applicable to transmitter housing

X1

Color Baked Epoxy

Item 5: Lightning Protection Device

no


A

Transmitter power supply voltage: 10.5-32V DC (intrinsic safety type: 10.5-30V DC, fieldbus communication: 9-32V DC) Maximum allowable current: 6000A (1X4 μ S), repeated 1000A (1X40 μ S) 100 times

Item 6: Prohibition of oil treatment or prohibition of water and oil treatment

no


K1

Degreasing and cleaning treatment

K5

Defatting, washing and drying treatment

Item 7: Calibration Unit

no


D1

P correction (in psi units)

D3

Bar calibration (bar unit)

D4

M correction (in kgf/cm2 units)

Item 8: Sealing treatment of SUS 630 nut

no


Y

Surface coating (liquid silicone rubber) on SUS630 bolts and nuts for the flange of the forbidden solid capacity chamber

Item 9: Non serrated shape

no


Q

The flange sealing surface is not serrated (only ANSI flanges)

Item 10: Zero point compensation for liquid contact temperature

no


R

Compensation range: 80-300 ℃

Item 11: PVC (vinyl chloride) is not protected outside the capillary tube

no


V

The maximum operating temperature is 100 ℃, used in environments where the use of PVC (vinyl chloride) is prohibited

Item 12: Quick Response

no


F1

Refresh time: ≤ 0.125 seconds Response time (including minimum damping time constant): 0.2 seconds (including isolation diaphragm)

Item 13: When abnormal, set the output to low side or NAMUK NE43

no


C1

Low side: ≤ -5% (3.2mm)

C2

(NAMUK NE43) [Output signal 3.8-20.5mA] Fault alarm output -5%, below 3.2mA

C3

(NAMUK NE43) [Output signal 3.8-20.5mA] Fault alarm output 110%, above 21.6mA

Item 14: Stainless steel amplifier housing

no


E1

Amplifier casing material: SCS14A stainless steel (equivalent to SUS316L cast stainless steel or ASTM CF-8M)

Item 15: PID/LM Function

no


LC1

PID control function, Link Master function

Item 16: Gold plated film

no


A1

Gold plating on sealing membrane

Item 17: Accessory Manufacturing Certificate

no


M05

Flange and diaphragm seat

M06

Flange, diaphragm seat, base, insertion cylinder

Item 18: Proof of Stress Test/Leakage Test

no


T41

A membrane box JIS 10K test pressure: 2MPa {20kgf/cm2} test fluid: nitrogen (N2) retention time: 10 minutes

T42

A membrane box JIS 20K test pressure: 3MPa {30kgf/cm2} test fluid: nitrogen (N2) retention time: 10 minutes

T43

A membrane box JIS 40K test pressure: 3MPa {30kgf/cm2} test fluid: nitrogen (N2) retention time: 10 minutes

T46

A membrane box ANSI/JPI 150 test pressure: 3MPa {29.8kgf/cm2} test fluid: nitrogen (N2) retention time: 10 minutes

T47

A membrane box ANSI/JPI 300 test pressure: 3MPa {30kgf/cm2} test fluid: nitrogen (N2) retention time: 10 minutes

T31

B membrane box JIS 10K test pressure: 2MPa {20kgf/cm2} test fluid: nitrogen (N2) retention time: 10 minutes

T32

B membrane box JIS 20K test pressure: 5MPa {50kgf/cm2} test fluid: nitrogen (N2) retention time: 10 minutes

T34

B membrane box JIS 40K test pressure: 7MPa {70kgf/cm2} test fluid: nitrogen (N2) retention time: 10 minutes

T36

B membrane box ANSI/JPI 150 test pressure: 3MPa {29.8kgf/cm2} test fluid: nitrogen (N2) retention time: 10 minutes

T38

B membrane box ANSI/JPI 300 test pressure: 7MPa {70kgf/cm2} test fluid: nitrogen (N2) retention time: 10 minutes

Similar Product Recommend