1Dimensions of Wuhan walk-in constant temperature and humidity test room
Studio standard volume |
0.5 ,1.0,2,……25,30,40cubic meter |
temperature upper limit |
150℃ |
lower temperature limit |
-70℃ |
temperature deviation |
±2℃ |
temperature fluctuation |
±0.5℃ |
humidity range |
30-98%R.H |
Humidity error |
+2/-3%R.H |
warm-up time |
average 3℃/min |
cooling time |
average1℃/min |
cooling method |
water cooling |
control manufacture device |
Chinese LCD touch screen temperature controller,PIDcontrol algorithm |
User program capacity |
120group1200paragraph(120group,1200sections) |
Security protection measures |
Temperature upper limit control,Compressor overpressure,Compressor overload,Heater short circuit |
material quality |
SUS 304#Stainless steel plate or storage plate |
make, do, create, compose, write, etc. use |
Program controlled temperature and humidity |
electricity source |
380V, 50Hz ,three-phase five-wire |
Note: A walk-in constant temperature and humidity laboratory can also be designed according to customer requirements to ensure customer satisfaction | |
IIDimensions of Wuhan walk-in constant temperature and humidity test roomfeature
1. Setting method: Touch, click
2. Display mode: Color LCD dot matrix touch screen with Chinese display
3. Setting and display resolution: temperature (0.1 ℃); Time (1 minute)
4. Graphic display: fully display the set program curve.
5. Set parameter storage time: After fully charged, data can be saved for 5 years.
6. Number of programs: 1-499 (up to 499 programs).
7. Program segments: 1 to 64 segments per program; Can run by group connection.
8. Can automatically prompt users to correctly set temperature and time parameters.
9. There is a dedicated maintenance interface for debugging and maintaining equipment with program running maintenance functions.
10. It has the function of waiting for program execution.
11. Has program skip function.
12. Has program stop function.
13. There is a power outage recovery function.
14. Control mode: constant temperature, program.
15. It has the function of locking the operating interface. Recording function: It can record curves and experimental data within 100 days, and can query the temperature and humidity situation at every moment within 100 days in detail. It can be exported using USB 2.0, printed on a PC, and generate data reports (equivalent to the function of a paperless recorder). It has a self checking function for startup faults.

IIIWuhan walk-in constant temperature and humidity test roomCharging method of refrigerant:
There are two methods for fluorine filling in split refrigeration equipment: high-pressure liquid refrigerant filling and low-pressure gas refrigerant filling. The method of high-pressure liquid refrigerant filling is often used for refrigerant filling during the production process of refrigeration equipment or for refilling refrigerant after major repairs of large air conditioning units. The method of low-pressure gas refrigerant charging is often used for supplementary operations of large refrigeration equipment or refrigerant refilling of general household refrigeration equipment. The charging operation method for split type refrigeration equipment is as follows:
(1) Method of filling refrigerant from the high-pressure side of refrigeration equipment (weighing method)
(2) Method of filling refrigerant from the low-pressure side of refrigeration equipment
4Wuhan walk-in constant temperature and humidity test roomEnvironmental conditions:
Temperature: 5 ℃~+35 ℃ (average temperature within 24 hours ≤ 28 ℃);
Relative humidity: ≤ 85% RH;
Air pressure: 86kPa~106kPa
Power supply conditions: three-phase four wire+protective ground wire, voltage range: AC (380 ± 38) V
Allowable frequency fluctuation range: (50 ± 0.5) Hz
The grounding resistance of the protective grounding wire is less than 4 Ω;
Require users to configure corresponding capacity air or power switches for the equipment at the installation site, and this switch must be used independently for the equipment;
Water supply conditions: Purified water is required for equipment humidification.