KPa, MPa, kg/cm2, psi, mbar, bar, inHg, inH2O, mH2O, mmH2O, and mmHg Pressure resistance limit optional pressure unit display pressure bar chart display refresh * small/* large value switch test voltage output alarm output adjustable installation position menu lock RS232 interface network function universal pressure module (UPM) battery pressure interface packaging European standard SiCalPro software system requirements universal parameters storage temperature size weight mechanical vibration mechanical impact electrical environment options
| 2 times the working pressure [20000 psi (1400 bar) range pressure limit is 29000 psi (2000 bar)]. |
| KPa, MPa, kg/cm2, psi, mbar, bar, inHg, inH2O, mH2O, mmH2O, and mmHg |
Pressure reading: 5-digit number, letter height 0.5 in (12.7 mm) Percentage indication: 2.5 digits, letter height 0.25 in (6.35 mm) |
| In addition to digital pressure indication, the DPI 104 LCD also includes a circular bar chart with 20 sub segments, allowing users to quickly visually estimate the pressure output from 0 to 100% full-scale. The increment of the bar chart is 5% of the user's selection range. |
| 2 times per second |
| *Small/large pressure values can be displayed on DPI 104, which can be enabled or disabled, and users can also reset the display value. |
| DPI 104 has the function of switch testing, which can capture and display the pressure values when the external pressure switch is opened and closed. *The impedance of the large switch is 200 Ω. |
| An analog output signal of 0-5V can be programmed to be proportional to the pressure displayed on the monitor, or set to a constant value. The voltage output mode provides 0.1% FS accuracy for 50 mV~5 V. |
| The alarm output node is implemented through open leakage field-effect transistor (* high current 250mA, * high voltage 24VDC) |
| For added convenience, the casing of DPI 104 can rotate around the pressure pipe joint by 320 °, and the front disc can also be rotated to any position for the best viewing angle. |
| Can prevent unauthorized menu access and settings. |
| Provide serial communication function, data can be transmitted to PC through the optional serial communication cable (IA4090-2-V0). Using this communication cable, all menu commands and display data can be released through ASCII instruction set or the optional SiCal Pro software. |
| Supports up to 99 DPI 104 devices connected together through a daisy chain network connection |
| The UPM module can be connected to DPI 104 to improve accuracy and expand pressure range |
| DPI 104 is equipped with one 9V alkaline battery, model MN1604. If used for 1 hour a day, the battery life is approximately 6 months. To improve power supply performance, it is recommended to use 9V lithium batteries (not provided by GE). |
1/4 NPT or BSP external thread pressure interface, with a range of 10000 psi (700 bar) or less. 9/16 x 18 UNF external cone thread pressure interface, with a range of 15000 psi (1000 bar) and 20000 psi (1400 bar) |
Shell material ABS/PC plastic, sealing type 4/IP65 Media components: media compatible with stainless steel (316) and Inconel |
| CE marking |
| This software package allows users to remotely control DPI 104 through a visualization instrument panel on their computer screen. The verification data can be recorded, browsed, or printed as a verification certificate. |
| *Low requirements: Intel Pentium processor, Windows ® 95 or higher version |
| -4°F~ 158°F (-20°C ~ 70°C) |
Diameter: 3.74 "(95 mm), excluding pressure fittings Thickness: 2.17 "(55 mm) |
| Approximately 13 ounces (350 g) |
| Complies with Def Stan 66-31, 8.4 Cat III |
| Compliant with BS EN 61010:2001 |
Electromagnetic compatibility: BS EN61326-1:1998+A2:2001 Electrical safety: BS EN 61010:2001 Mechanical safety equipment orientation level: SEP |
DPI 104 with wire external power supply, part number 191-350 PC serial port connection cable, part number IA4090-V0SiCalPro software, part number 1S-SiCALPRO-DPI104 (including serial communication cable IA4090-V-0) 8-pin output slot for switch testing, voltage output, external power supply, alarm output, etc. (when customers are willing to use their own connectors and wires), part number 1S-04-0027 9/16 x 18 UNF to 3/8 BSP adapter, part number 182-190, for connecting PV 212 hand pump DPI 104 UPM power cord, part number IA4101-1-V0, used to connect the power supply to the UPM IDOS sensor. Option H is requested. Universal 12V external power supply, part number 191-129, used to power UPM IDOS remote sensors |
Pneumatic and Hydraulic Test Kit ComponentsDPI 104 is included as a standard component in these inspection and verification kits: Low pressure air pressure detection kit, air pressure testing kit, hydraulic testing kit, air pressure and hydraulic testing kit
| Includes: DPI 104; Range up to 30 psi (2 bar), PV210 low-pressure air pressure test pump, pressure pipe, fittings, sealing components, and portable case. |
| Includes: DPI 104; Range up to 300 psi (20 bar), PV211 pressure test pump, pressure pipe, fittings, sealing components, and portable case. |
| Includes: DPI 104; Range up to 15000 psi (1000 bar), PV212 pressure test pump, pressure pipe, fittings, sealing components, and portable case. |
Includes: DPI 104; Range up to 10000 psi (700 bar), PV411A pneumatic hydraulic test pump, hydraulic reservoir, pressure pipe, fittings, sealing components, and portable case.
|
|