Welcome Customer !

Membership

Help

Shandong Haihui Environmental Technology Co., Ltd
Custom manufacturer

Main Products:

ybzhan>Products

Shandong Haihui Environmental Technology Co., Ltd

  • E-mail

    marketing@hve-china.com

  • Phone

  • Address

    Building B9, North High tech Industrial Corridor, Dongwu Road and Fenhe Road, Dongying City, Shandong Province

Contact Now

Explosion proof VOCs online continuous monitoring system

NegotiableUpdate on 11/04
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Explosion proof VOCs Online Continuous Monitoring System/Overview: For key industries in explosion-proof areas, we have launched an explosion-proof VOCs online continuous monitoring system. The explosion-proof cabinet used is a positive pressure blowing explosion-proof cabinet, with certificate number CNEx16.0316 and explosion-proof mark ExdeibpzIIBT4Gc

Product Details

Explosion proof VOCs online continuous monitoring system/Overview

We have launched an explosion-proof VOCs online continuous monitoring system for key industries in explosion-proof areas. The explosion-proof cabinet used is a positive pressure blowing explosion-proof cabinet, with certificate number CNEx16.0316 and explosion-proof mark ExdeibpzIIBT4Gc.

Explosion proof VOCs online continuous monitoring system/principle

VOCs连续在线监测系统

The explosion-proof VOCs online continuous monitoring instrument sampling system extracts the measured gas from the sampling point, passes through the high-temperature sampling probe for dust removal, and enters the online gas chromatograph through the high-temperature tracing pipeline. The chromatograph is equipped with a built-in heating box, which keeps all the pipelines through which the sample passes at high temperature. Then, a high-sensitivity hydrogen flame ionization detector (FID) is used to detect the sample, and finally, data collection, analysis, processing, transmission, and storage are automatically completed through workstation software.

Explosion proof VOCs online continuous monitoring system/feature

  • Adopting positive pressure blowing explosion-proof cabinet

    The VOCs online continuous monitoring instrument adopts a positive pressure blowing explosion-proof cabinet, which complies with the GB3836-2010 explosion-proof standard;

  • Using gas chromatography method

    The VOCs online continuous monitoring instrument adopts gas chromatography, which is an internationally recognized standard method for VOCs detection;

  • The preprocessing method has high reliability

    The pre-treatment method of the VOCs online continuous monitoring system meets the standards for the determination of fixed source exhaust gases in the United States, the European Union, and China, and has high reliability;

  • Adopting the total heat method

    Adopting the total heat method, the entire process from sampling to analysis is at high temperature without the need for water removal, effectively avoiding sample loss and ensuring accurate and reliable monitoring data;

  • Fast sampling response speed

    VOCs continuous monitoring instrument has a fast bypass flow path and fast sampling response speed;

  • Automatic blowing function

    The VOCs continuous monitoring system has an automatic blowing function, which can automatically remove dust on the surface of the filter element and extend the service life of the filter element;

  • Automatic calibration function

    VOCs online continuous monitoring instrument has automatic calibration function, no need for duty, and minimizes maintenance to the greatest extent possible;

  • Can monitor over a hundred types of organic waste gases

    The VOCs continuous monitoring system can monitor total hydrocarbons, non methane total hydrocarbons, and hundreds of organic waste gases, meeting the monitoring needs of different customers;

  • Adopting internationally advanced MFC

    Adopting internationally advanced MFC mass flow controller to control hydrogen, air, and carrier gas, the flow control accuracy can reach ± 0.05ml/min;

  • Optional configuration

    Standard configuration can be selected according to customer requirements;

Explosion proof VOCs online continuous monitoring system/parameter

project VOCs online continuous monitoring system parameters
Detection factor Total hydrocarbons, methane, non methane total hydrocarbons, benzene derivatives, alcohols, lipids, halogenated hydrocarbons, etc
range Methane 0-10/100/1000/10000/100000mg/m3(Optional);
Non methane total hydrocarbon methane 0-10/100/1000/10000/100000mg/m3
detector GC-FID Hydrogen Flame Ion Detector
detection limit Methane ≤ 0.1ppm, non methane ≤ 0.05ppm, benzene ≤ 0.03ppm
repetitiveness RSD≤2%.
Analysis cycle ≤3min
zero drift ≤±1%FS
range drift ≤±2%FS
Power supply <500W,220V AC/50Hz
working environment Temperature: (5-40) ℃, Humidity: (10% -90%) RH
Air source requirements Carrier gas: high-purity nitrogen or zero grade air (≥ 99.999%); Combustion gas: high-purity hydrogen gas (≥ 99.999%) Auxiliary gas: zero grade air (hydrocarbons<20ppb)
output 4-20mA、RS232/RS485、 Ethernet
size Height 1900 * Length 1200 * Width 800mm

Explosion proof VOCs online continuous monitoring system/application

The explosion-proof VOCs online continuous monitoring system can be used in explosion-proof areas such as petroleum and petrochemical, organic solvent manufacturing, electronics industry, spraying, medicine, automobile manufacturing, and rubber product printing.