We have launched an explosion-proof VOCs online continuous monitoring system for key industries in explosion-proof areas. The explosion-proof cabinet used is a positive pressure blowing explosion-proof cabinet, with certificate number CNEx16.0316 and explosion-proof mark ExdeibpzIIBT4Gc.
The explosion-proof VOCs online continuous monitoring instrument sampling system extracts the measured gas from the sampling point, passes through the high-temperature sampling probe for dust removal, and enters the online gas chromatograph through the high-temperature tracing pipeline. The chromatograph is equipped with a built-in heating box, which keeps all the pipelines through which the sample passes at high temperature. Then, a high-sensitivity hydrogen flame ionization detector (FID) is used to detect the sample, and finally, data collection, analysis, processing, transmission, and storage are automatically completed through workstation software.
Adopting positive pressure blowing explosion-proof cabinet
The VOCs online continuous monitoring instrument adopts a positive pressure blowing explosion-proof cabinet, which complies with the GB3836-2010 explosion-proof standard;
Using gas chromatography method
The VOCs online continuous monitoring instrument adopts gas chromatography, which is an internationally recognized standard method for VOCs detection;
The preprocessing method has high reliability
The pre-treatment method of the VOCs online continuous monitoring system meets the standards for the determination of fixed source exhaust gases in the United States, the European Union, and China, and has high reliability;
Adopting the total heat method
Adopting the total heat method, the entire process from sampling to analysis is at high temperature without the need for water removal, effectively avoiding sample loss and ensuring accurate and reliable monitoring data;
Fast sampling response speed
VOCs continuous monitoring instrument has a fast bypass flow path and fast sampling response speed;
Automatic blowing function
The VOCs continuous monitoring system has an automatic blowing function, which can automatically remove dust on the surface of the filter element and extend the service life of the filter element;
Automatic calibration function
VOCs online continuous monitoring instrument has automatic calibration function, no need for duty, and minimizes maintenance to the greatest extent possible;
Can monitor over a hundred types of organic waste gases
The VOCs continuous monitoring system can monitor total hydrocarbons, non methane total hydrocarbons, and hundreds of organic waste gases, meeting the monitoring needs of different customers;
Adopting internationally advanced MFC
Adopting internationally advanced MFC mass flow controller to control hydrogen, air, and carrier gas, the flow control accuracy can reach ± 0.05ml/min;
Optional configuration
Standard configuration can be selected according to customer requirements;
| project | VOCs online continuous monitoring system parameters |
| Detection factor | Total hydrocarbons, methane, non methane total hydrocarbons, benzene derivatives, alcohols, lipids, halogenated hydrocarbons, etc |
| range | Methane 0-10/100/1000/10000/100000mg/m3(Optional); Non methane total hydrocarbon methane 0-10/100/1000/10000/100000mg/m3 |
| detector | GC-FID Hydrogen Flame Ion Detector |
| detection limit | Methane ≤ 0.1ppm, non methane ≤ 0.05ppm, benzene ≤ 0.03ppm |
| repetitiveness | RSD≤2%. |
| Analysis cycle | ≤3min |
| zero drift | ≤±1%FS |
| range drift | ≤±2%FS |
| Power supply | <500W,220V AC/50Hz |
| working environment | Temperature: (5-40) ℃, Humidity: (10% -90%) RH |
| Air source requirements | Carrier gas: high-purity nitrogen or zero grade air (≥ 99.999%); Combustion gas: high-purity hydrogen gas (≥ 99.999%) Auxiliary gas: zero grade air (hydrocarbons<20ppb) |
| output | 4-20mA、RS232/RS485、 Ethernet |
| size | Height 1900 * Length 1200 * Width 800mm |