The FCC-1500H explosion-proof air sampler is an instrument used to collect gaseous or vapor samples from the atmosphere. It is suitable for sampling individual toxic substances and dust substances such as metals at designated locations for occupational hazards in the workplace; It is also applicable to routine and emergency monitoring of the environment, atmosphere, and public health
FCC-1500H Explosion proof Air SamplerIt is an instrument used to collect gaseous or vapor samples from the atmosphere, suitable for sampling individual toxic substances and dust substances such as metals at designated locations for occupational hazards in the workplace; It is also applicable to the environment and atmosphere,public healthRoutine monitoring and emergency monitoring.
1、 Application Fields
JJG956-2013air samplerVerification regulations
GBZ/T300-2017 Sampling Specification for Hazardous Substances in Workplace Air
IITechnical Parameter
1. Flow range:(10~200)mL/min,(20~500)mL/min,(50~1200)mL/min,(100-1500) mL/min, (100-3000) mL/min (flow rate optional)resolving power:1mL/min
2Constant current mode: Constant current and constant voltage(Accurate closed-loop control of flow is not affected by voltage fluctuations and air resistance changes, with continuous constant current and constant voltage sampling
3Flow calibration: an open backend calibration system where users can perform self calibration on the instrument.
4. Volume calculation: Automatic calculation of working condition volume and standard condition volume
5. Data storage: can store 99 sets of sampling data
6. Traffic volumeError: not exceeding±2.5%
7LCD Display Screen: Providedreal timeTraffic, running timeStandard condition volumeVarious information such as battery level, etc
8. Traffic stability: ≤ 5%
9. Traffic repeatability: ≤ 2%
10. Timing error: ≤± 0.1%
11. Timing range: 1min to 99h59min
12. Working temperature: -10~50 ℃
13. Relative humidity: ≤ 95% RH
14Ultra long battery life:>30 hours
15. Load capacity: ≥ 35000pa
16. Sampling modes: timed sampling, constant volume sampling, cyclic sampling
17. Host size: 122 * 60 * 72mm
18. Operating noise: < 60dB (A)
19. Host weight: 500g
20. Explosion proof type: intrinsic safety typeExibIIAT3Gb Explosion proof certificate number:SHExC20.1639
21Keyboard lock: Supports keyboard lock to prevent accidental operations from affecting the sampling process.
22. Temperature display: Yes; Atmospheric pressure display: Yes
Configuration List:
![]()