-
E-mail
nxyq530@163.com
-
Phone
13057528728
-
Address
Gaochun County, Nanjing City, Jiangsu Province
Nanjing Nuoxin Analytical Instrument Co., Ltd
nxyq530@163.com
13057528728
Gaochun County, Nanjing City, Jiangsu Province

The NX type furnace front carbon silicon analyzer adopts an anti electromagnetic interference, dust-proof, ultra-thin and portable design, which is easy to operate. Non professionals can quickly detect the carbon and silicon content of molten iron in front of the furnace after simple training.
One main feature: By using an improved evaluation method, important metallurgical parameters can be automatically controlled to compensate for the difficulty of accurately measuring non-metallic elements (C, Si) in "spectra" and the inability of conventional analytical instruments to meet the time requirements for rapid analysis in front of the furnace, meeting the quality control requirements of casting production. Two application functions: measuring C%, Si%, CEL, SC, △ T, △ TM, etc. for white cast iron and gray cast iron, and also calculating parameters such as RM, HB, Z/H, and MEC for non alloy cast iron.
Three technical parameters: Display digital height: 500mm Temperature compensation range: 0-15 ℃ Matching cup type: K-type Measurement accuracy: ± 1 ℃
Display temperature unit: Fahrenheit or Celsius temperature can be set. Maximum measurement time: 240 seconds
Four working procedures: Measurement status display: The instrument displays the "Ready" green light, "Measurement" yellow light, and "Completed" red light in a cyclic manner
Five power requirements: 98-242VAC, 50-60Hz, around 30VA
Instrument protection standard: IP65 standard protection
EMC standard: Implement EN50081-2 and EN50082-2 anti-interference standards
Six measurement content range:
Carbon equivalent CE measurement range 3.2-4.8% accuracy: ± 0.1%
Carbon content C: measurement range 2.8-4.2% accuracy: ± 0.04%
Silicon content Si: measurement range 0.9-3.0% accuracy: ± 0.1%