The GZX digital illumination incubator is a constant temperature cultivation device for the cultivation and breeding experiments of bacteria, molds, and microorganisms. It is particularly useful as an ideal equipment for scientific research and production in fields such as biotechnology, medical research, agricultural and forestry science, aquaculture, and animal husbandry

Product Details

GZX Digital Illumination IncubatorProduct Details
1、 Introduction:
It is a constant temperature cultivation device for the cultivation and breeding experiments of bacteria, molds, and microorganisms, and is particularly suitable for scientific research and production in fields such as biotechnology, medical research, agricultural and forestry science, aquaculture, and animal husbandry.
IIGZX Digital Illumination IncubatorProduct features:
1. It has super temperature and sensor abnormality protection function to ensure the safety of instruments and samples; Choosing a full spectrum plant growth lamp is beneficial for plant growth and improves disease resistance.
2. It has power-off memory and automatic power-off time compensation function; Constant temperature control system, fast response, high temperature control accuracy.
3. Microcomputer fully automatic control, touch switch, easy to operate; Programmable multi-stage control mode, temperature, humidity, light intensity, and time can be set separately during both day and night.
4. Air duct ventilation, with soft air velocity and uniform temperature in the working room; Adopting hollow reflective tempered laminated glass, it has good insulation performance and is aesthetically pleasing.
Optional fully enclosed opaque lampshade, optional studio power supply, disinfection device, etc.
Note: The B after each model represents two time periods, day and night; D represents 1-8 time periods. Among them, temperature, lighting, and time can all be set in multiple segments.
3、 Technical parameters:


model

GZX-150

GZX-250


Effective volume

150 liters

250 liters


Temperature range

5—50℃

5—50℃


Temperature Accuracy

1℃

1℃


compressor power

130W

185W


Power

150W

250W


Power supply

AC 220V 50HZ



Explanation: Refrigeration refers to the incubator being at an ambient temperature of 16-32 degrees Celsius.

(Light can be selected, for example: seed germination can be selected at 3500LX, and seedling growth can be selected at 7500LX or above.). All incubators in our company can increase the light intensity according to user requirements

4、 Instructions for use:

1. Connect the power supply, turn on the power switch, and the whole machine will be powered on. The power indicator light inside the switch will light up.

2. Please refer to the manual of the intelligent digital temperature controller for temperature control settings.

3. If lighting is required, turn on the lighting switch.

4. This incubator has a breakpoint protection function and a delay function of about one and a half minutes. It takes about one and a half minutes to restart the compressor after it stops.

5、 Maintenance and cultivation:

1. The outer shell of the incubator should be reliably grounded.

2. The incubator should be placed in a cool, dry, well ventilated place away from heat sources and direct sunlight. Place it steadily to prevent vibration and noise.

3. To ensure effective heat dissipation of the condenser, the distance between the condenser and the wall should be greater than 100mm. There should be a 50mm gap on the side of the box, and at least 300mm of space at the top of the box.

4. When handling, repairing, and maintaining the incubator, collision, shaking, and vibration should be avoided; The maximum inclination is less than 45 degrees.

5. The instrument suddenly doesn't work. Please check if the fuse tube (behind the box) is burnt out and check the power supply.

6. When cooling this incubator, it is not advisable to make the temperature difference between the incubator and the ambient temperature greater than 25 degrees.