Gas liquid pressure calibration table

NegotiableUpdate on 12/01
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The Gas Liquid Pressure Calibration Platform HS-YFT-2002L is an integrated pressure calibration equipment with pressure control fine-tuning and pressure output functions, as well as a pressure output of -0.095~60MPa. It is an economical pressure measuring instrument designed by our company based on more than ten years of experience in pressure instrument manufacturing and on-site application.
Product Details

Gas liquid pressure calibration table

High quality imported hydraulic pumps are selected as pressure generating components, supplemented by intricately designed fine-tuning, quick interface, and stabilizing valves. The advantages of simple operation, reasonable structure, and good sealing performance make it a wide range pressure generating system that can be widely used in industries such as metrology, scientific research, industrial and mining enterprises, and instrument manufacturing.

Gas liquid pressure calibration table

1 . Gas and liquid can be operated independently. The left side has an output gas pressure source of -0.095~0MPa, and the right side has an output liquid pressure source of 0~60MPa, making it convenient to carry and use.

2. Technical indicators

2.1Pressure range: -0.095-60MPa

2.2 Sensitivity adjustment: min 10 Pa

2.3 Dimensions: 400 × 400 × 160 (length × width × height)

2.4 Weight: 15kg

2.5 Medium: High voltage transformer oil, low voltage clean air

3. Usage operation

3.1. Forward detection

Unscrew the plugs on the connectors of output 1 and output 2, open the stabilizing valve, adjust the micro adjustment to the appropriate position (middle position), and close the air delivery valve. Gently operate the lever to direct the pressure liquid to the two pressure output ports. Reinstall the standard meter and the verified meter, operate the lever to increase the pressure to around the verification point, use fine-tuning to adjust to the verification point, and test point by point until the full range is reached.

3.2. Return Inspection

According to the regulations, after two readings at full range, slowly reduce the voltage to the calibration point with a fine adjustment, and test point by point until zero. When the adjustment cannot reduce the pressure, you can gently open the relief valve to reduce the pressure.

4. Precautions

4.1 High pressure liquid pressure pumps should be placed on a flat and easy to operate table to avoid use in environments with highly corrosive gases and dust particles.

4.2 The tested instrument should be properly cleaned to prevent stains from entering the detection circuit.

4.3 Output 1 and Output 2 connectors are quick couplings with sealed end faces,