1HSJS-IV frequency conversion anti-interference dielectric loss testerOverview
HSJS-IV frequency conversion anti-interferenceDielectric loss tester is a high-precision testing instrument used to measure the dielectric loss tangent and capacitance of various high-voltage power equipment in power plants, substations, and other fields or laboratories. The instrument is an integrated structure with a built-in dielectric loss testing bridge, a variable frequency voltage regulating power supply, a step-up transformer, and SF6 High stability standard capacitor. The high-voltage source for testing is generated by the inverter inside the instrument and is used for testing the test sample after being stepped up by a transformer. The frequency can be changed to 50Hz, 47.5Hz52.5Hz, 45Hz55Hz, 60Hz, 57.5Hz62.5Hz, 55Hz65Hz, using digital notch technology to avoid the interference of power frequency electric fields on testing, fundamentally solving the problem of accurate measurement under strong electric field interference. It is also applicable to situations where a generator is used for power supply testing after a complete power outage. This instrument is equipped with an insulating oil cup and a temperature control device to test the dielectric loss of insulating oil.
IIHSJS-IV frequency conversion anti-interference dielectric loss testerMain Features
1Large LCD Chinese display
Easy to operate, the instrument is equipped with a full touch LCD display screen and a super comprehensive touch operation interface. Each process is very clear and understandable, and operators do not need additional professional training to use it. With just a gentle click, the entire measurement process can be completed, making it an ideal intelligent dielectric loss measurement device at present.
2Massive storage of data
The instrument is equipped with a calendar chip and a large capacity memory, which can store 200 sets of data. It can save the test results in chronological order, view the historical records at any time, and print and output them.
3Scientific and advanced data management
Instrument data can be exported through a USB flash drive, and can be viewed and managed on any PC using our company's software.
4Multiple testing modes
The instrument is tested using internal high voltage and internal standard methods, including forward connection, reverse connection, and self-excited connection.
5CVT testing in one step
This instrument can also test the dielectric loss and capacitance of fully sealed CVTs (capacitive voltage transformers) C1 and C2.
6Measuring CVT without dismantling the high-voltage lead wire
The instrument can accurately measure the dielectric loss and capacitance values of the CVT without removing the high voltage leads of the CVT.
7Measurement of C0 using CVT reverse shielding method
The instrument can use reverse shielding method to measure the dielectric loss value and capacitance value of C0 at the upper end of CVT.
8High speed sampling signal
The inverters and sampling circuits inside the instrument are all digitally controlled, and the output voltage is continuously adjustable.
9Multiple protection, secure and reliable
The instrument is equipped with multiple protection measures such as input voltage fluctuation, high voltage current, output short circuit, power failure, overvoltage, overcurrent, temperature, etc., ensuring the safety and reliability of the instrument. The instrument also has a grounding detection function to ensure that ungrounded devices are not allowed to increase voltage.
IIIMain technical parameters
| 1 | Terms of Use | -15℃∽40℃ | RH<80% | ||
| 2 | Anti interference principle | Frequency conversion method | |||
| 3 | Power supply | AC 220V±10% | Allow generator | ||
| 4 | high voltage output | 0.5KV∽10KV | Every 0.1kV | ||
| Accuracy | 2% | ||||
| Zui high current | 200mA | ||||
| Capacity | 2000VA | ||||
| 5 | Self-excited power supply | AC 0V∽50V/15A | 45HZ/55HZ 47.5HZ/52.5HZ 55HZ/65HZ 57.5HZ/62.5HZ Automatic dual frequency conversion | ||
| 6 | resolution | tgδ: 0.001% | Cx: 0.001pF | ||
| 7 | Accuracy | △ tg δ: ± (reading * 1.0%+0.040%) | |||
| △ C x: ± (reading * 1.0%+1.00PF) | |||||
| 8 | measurement range | tgδ | unrestricted | ||
| C x | 15pF < Cx < 300nF | ||||
|
| 10KV | Cx < 40 nF | |||
|
| 5KV | Cx < 150 nF | |||
|
| 1KV | Cx < 300 nF | |||
| CVT testing | Cx < 300 nF | ||||
| 9 | External dimensions(host)(mm) | 350(L)×270(W)×270(H) | |||
| External dimensions(accessory box)(mm) | 350(L)×270(W)×160(H) | ||||
| 10 | Memory size | 200 sets support USB data storage | |||
| 11 | Weight (host) | 22.75Kg | |||