E-mail
2850687972@qq.com
-
Phone
13538469076
-
Address
Dongcheng Lixin Jinhui Industrial Park, Dongguan City, Guangdong Province
Guangdong Aisili Testing Instrument Co., Ltd
2850687972@qq.com
13538469076
Dongcheng Lixin Jinhui Industrial Park, Dongguan City, Guangdong Province
The high and low temperature constant temperature test chamber achieves precise constant temperature control within the set temperature range through the collaborative work of precision refrigeration system, heating system, and control system. Equipment selection should focus on core parameters such as temperature range, uniformity, and fluctuation, while also paying attention to the safety protection function and long-term operational stability of the equipment.
1、 Core technology system
1. Structural system
Studio material: SUS304 stainless steel plate, thickness 1.2mm
Shell material: cold-rolled steel plate with double-layer epoxy resin spray coating
Insulation layer: high-density glass fiber cotton, thickness 100mm
Sealing design: High and low temperature resistant silicone rubber double-layer sealing strip
Observation window: Three layer vacuum electric heating film anti frost glass
Test hole: 50mm diameter, with sealing plug
Cable interface: Waterproof gland head
2. Refrigeration system
Refrigeration method: mechanical compression refrigeration
Circulating system:
Single stage compression cycle (0 ℃~-40 ℃)
Cascade compression cycle (-40 ℃~-80 ℃)
Compressor: Semi closed piston compressor
Condenser: air-cooled finned heat exchanger
Evaporator: Copper tube aluminum fin heat exchanger
Throttle device: thermal expansion valve+solenoid valve
Refrigerant: R404A/R23 (cascade system)
3. Heating system
Heating element: Nickel chromium alloy electric heating wire
Installation method: Stainless steel protective sleeve
Control mode: SSR solid-state relay regulation
Safety protection: Fused over temperature protection
4. Air circulation system
Fan type: Multi blade centrifugal fan
Motor power: 90W
Blade diameter: 200mm
Wind speed: adjustable from 0.5-1.5m/s
5. Control system
Main controller: Japan U-Easy Control UMC1200
Display interface: 7-inch color touch screen
Temperature sensor: PT100 platinum resistor
Control mode: PID+SSR control
Program capacity: 120 sets of programs, with 100 segments per set
Communication interface: USB, RS-485
Data recording: Built in storage, supports USB drive export
2、 Technical parameters
1. Temperature parameters
Temperature range: -70 ℃ to+150 ℃
Temperature fluctuation: ± 0.5 ℃
Temperature uniformity: ± 2.0 ℃
Heating time:+25 ℃ →+150 ℃ ≤ 45 minutes
Cooling time:+25 ℃ → -70 ℃ ≤ 90min
2. Structural parameters
Inner box size: 400 × 400 × 400mm (64L)
Outer box size: 950 × 1050 × 1550mm
Power supply voltage: AC220V ± 10% 50Hz
Maximum power: 4.5kW
Weight: Approximately 280kg
3、 Performance indicators
1. Temperature performance
Stability: Continuous operation for 24 hours, temperature fluctuation ≤ 0.5 ℃
Uniformity: 9-point test in workspace, temperature difference ≤ 2.0 ℃
Accuracy: The deviation between the displayed value and the measured value is ≤± 1.0 ℃
2. Reliability indicators
Average time: ≥ 5000 hours
Compressor lifespan: ≥ 50000 hours
Continuous operation capability: ≥ 720 hours
3. Security protection system
Electrical safety:
Leakage protection switch
Overcurrent protector
short circuit protection
Equipment safety:
Compressor overheat protection
Compressor overcurrent protection
Fan overload protection
Cooling fan interlock protection
Sample safety:
Independent over temperature protector
Temperature upper limit protection
Temperature lower limit protection
4、 Manufacturing and Inspection Standards
1. Design standards
GB/T 10592-2008 Technical Conditions for High and Low Temperature Test Chambers
IEC 60068-2-1 Test A: Low Temperature
IEC 60068-2-2 Test B: High Temperature
2. Inspection items
Insulation resistance test: ≥ 2M Ω
Voltage endurance test: 1500V/10mA/60s
Leakage current test: ≤ 3.5mA
Grounding resistance test: ≤ 0.1 Ω
Temperature performance test (empty/full load)
Security protection function verification
5、 Environmental requirements for use
Environmental temperature:+5 ℃~+35 ℃
Environmental humidity: ≤ 85% RH
Power supply fluctuation: ± 10% of rated voltage
Ventilation requirement: A space of ≥ 500mm around the equipment
Ground bearing capacity: ≥ 500kg/m?