Technical Specification for TLP700L-4 High and Low Temperature Test Chamber (Vertical Structure, Non standard Customized Model) 1. Main Technical Parameters: 1.1 Temperature Range: -40~130 ℃
High and low temperature test chamber (vertical structure, non-standard customized model)
TLP700L-4 Technical Specification
1. Main technical parameters:
1.1 Temperature range: -40~130 ℃;
1.2 Volatility: ≤ 0.5 ℃;
1.3 Temperature deviation: ≤± 2 ℃;
1.4 Heating time: Room temperature -+130 ℃ ≤ 60 minutes (no load);
1.5 Cooling time: Room temperature -20 ℃ ≤ 50min (with a heat load of 500W, which is a commonly used temperature point);
Room temperature -40 ℃ ≤ 70min(Whole process average, no load, ambient temperature 25 ℃);
1.6 Studio dimensions: D600 x W1800 x H650, mm, depth x width x height;
1.7 External dimensions: D1100 × W2000 × H1800, mm, depth × width × height;
1.8 Overall power: 10KW;
1.9 Installation of power supply: 380V+19V/-38V; 50 Hz;3P+N+G;
2. Air conditioning and refrigeration system:
2.1 Air conditioning mode: forced ventilation, internal circulation balance, temperature and humidity control;
2.2 Air circulation: large capacity axial flow fan;
2.3 Heating method: Nickel chromium electric heater;
2.4 Refrigeration System: Fully enclosed French Taikang compressor, compressor power: 6P; using single-stage refrigeration method;
2.5 Cooling method: air-cooled condenser;
2.6 Refrigerant: R404A;
2.7 Control cooling components: expansion valve, filter, solenoid valve;
3Measurement and Control System
3.1 Temperature measurement: Pt100 platinum resistor;
3.2 Control system: 5.7 "touch screen temperature control, LCD display and operation, South Korean" three yuan ", model: TEMI880;
3.3 Operation mode: Constant operation, program operation.
3.4 Main electrical appliances: using Tai'an and other electrical components
4. Safety protection measures: Safety Device
4.1 Whole machine power leakage protection;
4.2 Heater fuse protection;
4.3 Overtemperature protection: an overtemperature protection system independent of the studio main controller;
4.2 Overload of blower motor;
4.3 Refrigeration machine overpressure protection;
4.4 Refrigeration machine overload protection;
5. Test methods that the equipment meets:
5.1 GB/T2423 High temperature test, low temperature test, high and low temperature cycle test;
6. Environmental conditions for using the device:
6.1 Environmental temperature: 5~35 ℃;
6.2 Relative humidity: ≤ 85% R.H;
6.3 Atmospheric pressure: 86~106 kPa;
6.4 Installation of power supply: 380V, 50HZ;
7. Chamber Structure and Material Description: Chamber
7.1 Box material: The shell is mostly made of 0.8mm thick brushed stainless steel, and the reinforced area and load-bearing parts are made of 1.0mm to 2.0mm brushed stainless steel;
7.2 Interior material: SUS304 high-quality stainless steel plate is used;
7.3 Thermal insulation material: rigid polyurethane foam;
7.4 Observation window: electrically heated frost resistant insulated glass (size:300×450);
7.5 Cable hole: ф 50, left side, quantity 1;
7.6 Working lighting: 12W;
7.7 Control box: placed on top of the test box;
7.8 Test box door: Double open door, equipped with heavy-duty door hinges;
7.9 Door lock: Explosion proof door lock (commonly known as container lock);
7.10 Casters: Equipped with "Global" brand casters, bearing type 3 "polyurethane wheels, including 2 casters with brake devices.

Shenzhen Fuyida Instrument Co., Ltd. www.fuyidatest.com