E-mail
nai17#foxmail.com(#换@)
-
Phone
13916283642
-
Address
Building 1, No. 139 Rongmei Road, Songjiang District, Shanghai
Shanghai Naiai Precision Instrument Co., Ltd. (Headquarters)
nai17#foxmail.com(#换@)
13916283642
Building 1, No. 139 Rongmei Road, Songjiang District, Shanghai

The specifications of the infrared high-throughput intelligent microwave digestion instrument include: 6, 8, 10, 12, 16, 18, 20, 24, 40, and 44 positions. The microwave digestion instrument adopts microwave non pulse continuous automatic frequency conversion control, which extends the service life of the instrument and the uniformity of electromagnetic waves. The chamber is specially made of large volume stainless steel chamber material, with a self-locking buffer explosion-proof furnace door. When the reaction is abnormal, the buffer structure ensures the personal safety of the operator and the integrity of the furnace door structure. The furnace door and chamber are tightly combined, and microwave leakage meets national standards. The instrument adopts a temperature and pressure dual control system to control the pressure and temperature of the digestion experiment, and displays them in real time. 360 ° continuous rotation in the same direction, uniform microwave, ensuring the same microwave environment for each sample and improving the consistency of experimental results. When the pressure inside the tank exceeds the set protection value, the microwave will automatically stop heating. The safety explosion-proof film has dual safety functions, and the microwave digestion instrument provides a fast, safe, and automated solution for samples, accelerating the speed of sample digestion reaction under high pressure conditions. It is widely used in fields such as food, environmental protection, disease control, quality supervision, commodity inspection, and scientific research institutes.
Application scope
Widely used in fields such as food, environmental protection, disease control, quality supervision, product inspection, and research institutes.
Technical Parameter
| PRODUCT MODEL | NAI-WB44 | NAI-WB40 | NAI-WB24/20/18/16/12/10/8 | NAI-WB6 |
| Number of reaction tanks (set) | 44 | 40 | 24/20/18/16/12/10/8 | 6 |
| Power requirements | 220 VAC 50 Hz 15A | |||
| microwave source | Dual magnetron, professional microwave source, industrial grade microwave power supply | |||
| Microwave output power | 0~2200W (power adjusted according to configuration) | |||
| Microwave output characteristics | Microwave non pulse continuous automatic frequency conversion control, 0-100% automatic output | |||
| Microwave uniformity | Vertical bidirectional waveguide design, 3D output technology, matching harmonic power frequency to achieve high field uniformity | |||
| Microwave cavity | 66L, 316 grade all stainless steel chamber, 6-layer anti-corrosion and high-temperature resistant Teflon coating | |||
| Explosion proof furnace door | Self locking buffer explosion-proof furnace door can automatically release lateral pressure impact in advance when danger occurs, ensuring the personal safety of operators and the integrity of the furnace door structure | |||
| Exhaust and cooling system | The furnace chamber is equipped with a high-power exhaust system, and various reactions can be carried out continuously for a long time in a ventilated, safe, and easily observable environment. The furnace chamber ventilation adopts acid resistant and high air volume centrifugal fans, with an exhaust air volume of not less than 5m3/min; The furnace chamber has air cooling function, which continuously cools the reaction tank and displays temperature and pressure in real-time. | |||
| Rotary design | 360 ° continuous rotation in the same direction, uniform microwave, ensuring the same microwave environment for each sample, improving the consistency of experimental results, and distinguishing accessory damage caused by back and forth rotation | |||
| control mode | Adopting Siemens PLC from Germany | |||
| 8-inch touch screen | 8-inch TFT-LCD (800X480 color) large screen display, remote direct reading of reaction process, real-time display of closed reaction tank temperature and pressure, and real-time display of temperature pressure curve; Operating language: Simplified Chinese | |||
| Automatic protection function | Equipped with a fault self checking system; When one of the temperature and pressure control systems fails, the instrument will automatically cut off microwave transmission and sound an alarm | |||
| Acid remover | Equipped with two 20 seater acid washers | Equipped with a 20 seater acid washer | Equipped with a 6-position acid washer | |
| Control system parameters | ||||
| Temperature controller | Adopting non-contact temperature control technology, precise and error free temperature control, using high-precision long wave infrared sensors; Real time detection, control, and display of temperature and curve inside the microwave digestion reaction tank; Dual infrared full tank temperature control method. (The selection of infrared temperature control can be based on the number of reaction tanks to choose single/double infrared temperature control. Please refer to the instruction manual for details.) | |||
| Temperature control range | 0~350℃; Temperature control accuracy: ± 0.3 ℃; Actual operating temperature below 300 ℃ (adjust the maximum temperature according to the sample) | |||
| Temperature control capability | Rate heating and time heating functions. pid control | |||
| Pressure control system | Adopting non-contact pressure control method, precise and error free pressure control, and automatic pressure relief when the pressure exceeds the limit. Real time detection and display of pressure and curve inside the microwave digestion reaction tank | |||
| Pressure control range | Range: 0~10MPa (any pressure range from 0~15MPa can be selected) | |||
| pressure protect | Automatic adjustment/stop of microwave emission and automatic alarm for overpressure | |||
| Reaction tank parameters | ||||
| Temperature and pressure | Maximum temperature ≥ 350 ℃, maximum pressure ≥ 1500psi; Use temperature below 300 ℃. | |||
| Inner tank material | Polytetrafluoroethylene material; Imported modified polytetrafluoroethylene. TFM | |||
| Explosion proof outer sleeve material | High strength imported aerospace material PEEK | |||
| Internal tank reaction volume | Choose between 70ml and 100ml (standard 70ml) | |||
Detail shooting