Hot ball anemometer/QDF-6 thermoelectric anemometer The QDF-6 series thermal anemometer anemometer is a portable, instant display instrument with the basic function of measuring low wind speeds. It has the characteristics of intelligence and humanization, and is easy to use and operate. It has a wide range of applications in heating, ventilation, air conditioning, environmental protection, energy conservation monitoring, meteorology, agriculture, refrigeration, drying, labor hygiene investigation, and other fields
Hot ball anemometer/QDF-6 thermoelectric anemometer QDF-6Technical parameters of hot bulb anemometer:
Wind speed measurement range:0.05~30.0m/s;
Wind speed measurement accuracy:0.05~5.00m/s : ±(4%U+0.2)m/s;
5.0~30.0m/s : ±(4%U+0.3)m/s.
Additional error: The direction deviation of the measuring head is±15°When within, the additional error of the displayed wind speed value shall not exceed±5%UAmong themUTo measure the wind speed (same as above).
Response time: The response time of the probe is not greater than 3 In seconds.
Minimum resolution of wind speed:0.01m/s;
Airflow accuracy: Reference wind speed value
The front end of the probe can be bent for easy measurement;
Usage environment: ambient temperature5~40℃, relative humidity not exceeding90%RHIn clean air;
power supply 4section5Alkaline battery;
Battery life: continuous measurement10More than hours.
6 External dimensions and weight
6.1 External dimensions:187 ×84 ×46 mm
6.2 Measuring rod size:The longest:1025mmThe shortest:255mm
Probe diameter:φ11mm
6.3 Probe lead length: 3rice
6.4 weight: ≯600g(Contains battery)
Usage environment: ambient temperature5~40℃, relative humidity not exceeding90%RHIn clean air;
power supply 4section5Alkaline battery;
Battery life: continuous measurement10More than hours.
6 External dimensions and weight
6.1 External dimensions:187 ×84 ×46 mm
6.2 Measuring rod size:The longest:1025mmThe shortest:255mm
Probe diameter:φ11mm
6.3 Probe lead length: 3rice
6.4 weight: ≯600g(Contains battery)