Room 1001, Building 6, Xiexin Center, No. 19 Hefeng Road, Xinwu District, Wuxi City
Wuxi Huikete Measurement and Control Technology Co., Ltd
hkt@huikete.com.cn
13921186818
Room 1001, Building 6, Xiexin Center, No. 19 Hefeng Road, Xinwu District, Wuxi City
Hydrogen peroxide environmental temperature and humidity dew point probe
Probe used in hydrogen peroxide environment: sterilized with hydrogen peroxide. Hydrogen peroxide (H2O2) is used for disinfecting cleanrooms, incubators, and other equipment. In this process, the air humidity is saturated through spray, resulting in the formation of micro gel on all object surfaces. Hydrogen peroxide will kill all microorganisms. Then H 2 O 2 is mechanically recovered or naturally decomposed into harmless H 2 O or O 2 within 48 hours. The humidity level during the application process is crucial for the effectiveness of the process.
Standard humidity sensors are difficult to accurately measure in high concentration H2O2 environments. Hydrogen peroxide will occupy the micropores on the surface of the sensor to affect water vapor measurement. That's why ROTRONIC has developed a special sensor called HYGROMER ® HH-1。 Compared to standard sensors, the lifespan of sensors is greatly extended. Please refer to the sensor data sheet for the exact technical data of the sensor.
Technical parameters of hydrogen peroxide environmental temperature and humidity dew point probe:
| precision | ± 0.8% RH, ± 0.1 ° C, at 10... 30 ° C |
|---|---|
| Probe type | Temperature and humidity probe for hydrogen peroxide application (H2O2) |
| Integrated probe measurement range | -50…100 °C / 0…100 %rh |
| Analog output accuracy | ±1 mV |
| power supply | 3.3…5 VDC |
| sensor | Hygromer HH-1 |
| temperature sensor | PT100 1/3 DIN Class B |
| Filter type | none |
| Response time T 63 | <15 sec. |
| calculated value | Dew point/frost point |
| Anti H2O2 | 880 ppm / 1200 mg/m3 |
| current consumption | 4.5 mA, 3.3 VDC power supply |
| Audit Tracking/Electronic Records | FDA 21 CFR Part 11 and GAMP compatible |
| Analog output signal (standard) | 0…1 V = 0...100 %RH 0…1 V = -40...60 °C (default setting, freely scalable via HW4 software and AC3001 cable) |
| communication interface | UART |
| housing material | Stainless steel 1.4301 |
| IP protection rating | IP65 |
| CE/EMC certification | EMC Directive 2007/108/EC |
| Storage and Transportation Conditions | -50…100 °C |