E-mail
info#shkkgg.com
-
Phone
13371864393
-
Address
2086 Guyang North Road, Songjiang District, Shanghai
Guanhe Instrument Equipment (Shanghai) Co., Ltd
info#shkkgg.com
13371864393
2086 Guyang North Road, Songjiang District, Shanghai

Inductively coupled plasma emission spectrometer ICP-6800S (petrochemical specific machine)
Product Introduction
ICP-6800SThe petrochemical inductively coupled plasma emission spectrometer is an instrument specifically designed for the petrochemical industry, which can be used to determine the concentration of pollutants in crude oil30Multiple elements, mainly includingFe.Na.Mg.Ni.V.Ca.Pb.Mo.Mn.Cr.Ca.Ba.AsCan be used to measure iron in gasoline.manganese.lead.Silicon; Can be used as an additive in lubricating oil.Determination of wear elements; Can be used for multi-element determination in base oil; Can be used for the determination of sodium element in formaldehyde; Determination of oilfield tracer.
work environment
content Scope of adaptation
Storage and transportation temperature 15℃-25℃
Relative humidity during storage and transportation ≤70%
atmospheric pressure86-106 kPa
Power adaptability 220±10v 50-60MHz
Working humidity ≤70%
operation temperature 15℃-30℃
TECHNICAL INDEX
Technical indicators of solid-state power supply
Circuit type: Inductive feedback self-excited oscillation circuit, coaxial cable output, matching tuning, power feedback closed-loop automatic control
Working frequency:40.68MHz±0.05%
Frequency stability: <0.1%
Output power:800W—1200W
Output power stability:<0.3%
Electromagnetic field leakage radiation intensity: distance from the chassis30cmElectric field strengthE:<2V/m
Technical indicators of injection system
Output the inner diameter of the working coil25mm
Rectangular tube, three concentric shape, outer diameter20mmThe quartz rectangular tube
Outer diameter of coaxial spray6mm
Outer diameter of twin tube fog chamber34mm
Specification of argon flowmeter and carrier gas pressure gauge
1.Plasma gas flowmeter(100-1000)L/h (1.6-16L/min)
2.Auxiliary gas flow meter(10-100)L/h (0.16-1.66L/min)
3.Carrier gas flowmeter(10-100)L/h (0.16-1.66L/min)
4.Oxygen flow meter(0-100)ml/min (30-50ml/min)
5.Carrier gas stabilizing valve(0-0.4MPa)
6.Cooling circulating water: water temperature20-25℃ internet traffic>5L/minwater pressure>0.1MPa
Technical specifications of monochromator
Optical path:Czerny-Turner
Focal length:1000mm
Grating specifications: ion etched holographic grating, engraved line density4320line/mm
Reciprocal of linear dispersion rate:0.26nm/mm
Resolution: ≤0.005nm(4320Engraved lines)
Scanning wavelength range: 4320line/mmScanning wavelength range:180—435 nm
Stepper motor drives Z small step pitch:0.0006 nm
Exit slit:10μmIncident slit:8μm
Technical specifications of photoelectric converters
Specification of photomultiplier tube:R293
Negative high voltage of photomultiplier tube:0-1000VStability<0.05%
Fundamentals of Refrigeration System Technology
Refrigeration type swirl fog chamber: using imported silicone gel collection technology can quickly reduce the temperature to below zero20linear measure
Technical specifications of the whole machine
Scanning wavelength range:185nm~435nm(4320/mmGrating)
Repeatability: (i.e. short-term stability) relative standard deviationRSD≤1.5%
Stability: relative standard deviationRSD≤2%
Detection limit(mg/L):
Direct injection measurement of iron, manganese, lead, and silicon in gasoline analysis results:
element |
Iron(Fe) |
Manganese(Mn) |
Lead(Pb) |
Silicon(Si) |
measured value/(mg/L) |
3.6 |
17.5 |
1.8 |
2.4 |
RSD/% |
0.56 |
0.23 |
1.38 |
0.92 |
Labeling and recovery rate:
Serial Number |
the adding standard matter amount (mg/L) |
Iron(Fe) |
Manganese(Mn) |
Lead(Pb) |
Silicon(Si) |
||||
Measurement value(mg/L) |
rate of recovery /% |
Measurement value(mg/L) |
rate of recovery/% |
Measurement value(mg/L) |
rate of recovery /% |
Measurement value(mg/L) |
rate of recovery /% |
||
1 |
0.45 |
0.436 |
96.9 |
0.468 |
104.0 |
0.432 |
96.0 |
0.465 |
103.3 |
2 |
0.9 |
0.887 |
98.6 |
0.913 |
101.4 |
0.868 |
96.4 |
0.919 |
102.1 |