E-mail
2316712822@qq.com
-
Phone
17701823518
-
Address
Room 711, Xinyuan Business Building, 155 Xinyuan Road, Jiading District, Shanghai
Shanghai Xinsong Industrial Co., Ltd
2316712822@qq.com
17701823518
Room 711, Xinyuan Business Building, 155 Xinyuan Road, Jiading District, Shanghai
Jenco 6360 portable multi parameter water quality analyzer
Characteristics of 6360 portable multi parameter water quality analyzer
Measurable pH, ORP, EC, Salt, TDS, Temperature
Microcomputer design with RS-232C communication interface (can be converted to USB)
Automatic/manual temperature compensation
Auto Power Off
Jenco 6360 portable multi parameter water quality analyzer
Can connect 2-ring and 4-ring conductivity electrodes
Fifty sets of test data memory storage
AC/DC dual-purpose power supply
Jenco 6360 portable multi parameter water quality analyzer
Specification parameters
display |
range |
resolution |
linear measure |
pH |
-2.00~16.00pH |
0.01pH |
± 0.1% pH ± 1 word |
ORP |
-1999~1999mV |
1mV |
± 0.1% ± 1 word |
conductivity |
-0.00uS/cm~200.0mS/cm |
0.01/0.1 |
± 0.5%FS |
TDS |
0.00ppm~200.0ppt |
0.01/0.1 |
± 0.5%FS |
salinity |
0.0~80.0ppt |
0.1 |
± 1%FS |
temperature |
-5.0~100.0℃ |
0.1℃ |
± 0.3 ℃ ± 1 word |
Performance of 6360 portable multi parameter water quality analyzer
Types of electrode constants |
0.475/0.1/cm-1 |
Conductivity temperature compensation range |
Automatic -5.0~100.0 ℃ |
reference temperature |
15.0℃~25.0℃ |
temperature coefficient |
0.00~4.00% adjustable, default value is 1.91% |
TDS coefficient |
Adjustable from 0.30 to 1.00, default value 0.65 |
PH correction |
2-point automatic calibration (optional for USA and NIST groups) |
Conductivity correction |
Multi point manual calibration |
Salinity correction |
Multi point manual calibration (China's standard seawater salinity includes 5ppt, 20ppt, 30ppt, 35ppt, and 40ppt) |
Test data storage |
50 groups |
communication |
RS-232C (USB convertible) |
Automatic shutdown |
10 minutes without operation |
power supply |
AC power converter (output: DC9V) or 6 1.5 volt 7 # batteries |
work environment |
0~50 ℃, relative humidity<90% |
size |
241×86×72mm |
weight |
About 470g (excluding battery) |
Jenco 6360 portable multi parameter water quality analyzer standard configuration
6360 |
Host, four ring electrode (K=0.475cm-1), 6005P, AC power converter (output: DC9V) or 6 1.5 volt 7 # batteries, suitcase, software CD |
Jenco 6360 Portable Multi Parameter Water Quality Analyzer Accessories Selection
600S-ORP/600L-ORP |
Plastic shell/glass shell ORP composite electrode (redox potential measurement) |
GB700E |
Filler glass pH composite electrode (conventional measurement) |
GB710E |
Composite electrode (glass), viscous or oily object, strong acid and strong alkali |
600P |
Composite plastic shell pH composite electrode (without temperature compensation, conventional measurement) |
600C |
Pure water glass pH composite electrode (for measuring pure water and ultrapure water) |
6230AST |
Stainless steel temperature probe |