Shanghai instrument differential pressure transmitters are mainly used to measure the differential pressure, pressure, liquid level, and density of liquids, gases, or vapors, and then convert them into 4-20mA.DC signal outputs.
Product Details

Usage Instructions

Shanghai instrument differential pressure transmitterMainly used to measure the differential pressure, pressure, liquid level, and density of liquids, gases, or vapors, and then convert them into 4-20mA.DC signal outputs.
Differential pressure transmitters include three types: DP type (base type), HP type (high static pressure), and DR type (micro differential pressure). The above three types of transmitters combined with an intelligent amplifier board can form an intelligent type, which can communicate with each other through a handheld device that complies with the HART protocol for setting and monitoring.
The HP type high static pressure transmitter is suitable for high-pressure equipment in petrochemical and power plants. Due to its high pressure resistance of up to 32MPa and the ability to withstand unidirectional overload pressure without damage, it can be reliably used for differential pressure measurement in high-pressure systems.
The DR type can convert small differential pressures into 4-20mA DC signals. It uses specially designed amplification circuit boards and special temperature compensation processes, ensuring stable and reliable performance. It is an ideal product for detecting micro pressure, primary and secondary air flow, and airflow loss in furnace combustion chambers.

LED-1151 series code Transmitter selection code
Measurement type DR hhp
2 0 -0.125~1.5
Output signal type F 4-20mA.DC analog amplifier, linear output
S 4-20mA.DC intelligent amplifier, using HART protocol communication


Liquid receiving material
Container and connector Exhaust/Drain Valve


Isolation membrane

12 carbon steel 316 steel 316 steel
22 316 steel 316 steel 316 steel
Type of filling fluid none
B
silicone oil

Allowable static pressure value
A 2
B 5
display mode M1 Linear indicator with 0-100% scale

M2 Square root indicator scale 0-10
M4 Digital display indicator with 0-100% scale
Bracket form B1 Pipe installation bending bracket (installing pipe diameter 60)
B2 Board mounted bending bracket
B3 Pipe installation bending bracket (installing pipe diameter 60)
Location of exhaust/drain valve D1 The exhaust/drain valve is installed on the upper part of the side of the chamber
D2 The exhaust/drain valve is installed at the lower part of the side of the chamber
flameproof type d Explosion proof type, explosion-proof mark Exd Ⅱ BT4-6

LED-1151 DR 2 F 12 A M1 B2 Example of Selection for 0-0.4KPa