Portable multimeter calibrator | Portable multimeter calibrator | Portable multifunctional calibrator | Portable multimeter calibrator | Portable multimeter calibrator | Portable multimeter calibrator | Portable multimeter calibrator | Portable multimeter calibrator | Portable multimeter calibrator | Portable calibrator | Portable calibrator 1. Purpose and features: HG5610A multifunctional calibrator (multimeter calibrator) is suitable for verifying and calibrating various current and voltage meters below level 0.2
portableThree meter calibrator|portableThree meter calibrator|portableMulti functional calibrator|PortableTriple meter calibrator|PortableMulti purpose meter calibrator|PortableThree purpose meter calibrator|PortableMultimeter calibrator|PortableMulti meter calibrator|Portable calibrator
1Purpose and Characteristics:
HG5610AMulti functional calibrator(Three purpose meter calibrator)Suitable for calibration and verification of various types0.2Current and voltage meters below level. It can also be used as a high stability testing power supply, in conjunction with high-level standard tables, for calibration0.2Current and voltage meters above level.
2Main functions:
2.1、 5 1/2bitLEDDigital display output quantity.
2.2The output range of AC and DC currents is:0~50uA~5mA~50mA~500mA~2A~10A
2.3The output range of AC and DC voltage is:0~500mV~5V~50V~250V~500V~1000V
2.4There are three communication frequencies available for selection:50HzThe60HzThe400Hz
2.5Output overload can be automatically protected and manually reset.
2.6Each voltage and current level can be overloaded10%.
2.7Clamp gauge measurement: equipped with our standard coil, it can measureAC 0~750A(Some tables can be measured)1000AAbove);DC 0~1000A Error ±0.3%.
3Technical performance
3.1 Stability: DC<Fullrangeof0.01%/2minute
AC<Fullrangeof0.03%/2minute
3.2Communication distortion: <0.5%
3.3DC ripple factor:<0.1%
3.4Output frequency and accuracy:50Hz 60Hz 400Hz ﹤1%
3.5Range and accuracy of output voltage and current(23℃±2℃,The output value is greater than10%range)
|
Output project
|
output range
|
rated output
|
maximum output
|
accuracy
|
|
|
AC and DC
electricity press
|
0~500mV
|
10mA
|
30mA
|
DC: ± (reading)0.02%+fullrangeof0.03%)
AC: ± (reading)0.03%+fullrange of0.05%)
|
|
|
0~5V
|
20mA
|
60mA
|
|
|
0~50V
|
40mA
|
100mA
|
|
|
0~250V
|
20mA
|
60mA
|
|
|
0~500V
|
10mA
|
30mA
|
|
|
AC and DC
electricity flow
|
0~50uA~5mA~50mA
|
DC:3V
AC:36V
|
|
DC: ± (reading)0.02%+fullrangeof0.03%)
AC: ± (reading)0.03%+fullrangeof0.05%)
2mAThe following does not assess accuracy
|
|
|
|
|
0~500mA
|
DC:3V
AC:12V
|
|
|
|
|
0~2A~10A
|
2V
|
|
|
|
mark accurate
electricity block
|
10、50、100The
500(Ω)
1KThe50KThe100KThe500KThe
1MThe10MThe20MThe
|
0.25W
|
|
±0.3%+20mΩ
|
|
3.6Power consumption: AC power supply voltage220V±10%, frequency50Hz±1HzPower consumption<180VA
3.7Working environment: Temperature of the working environment5℃~35℃, relative humidity not exceeding80%
3.8Working hours: continuous
3.9External dimensions:135×400×360mm3
3.10Weight: Approximately12kg
[Keywords: portable multimeter calibrator | portable multimeter calibrator | portable multifunctional calibrator | portable multimeter calibrator | portable multimeter calibrator | portable multimeter calibrator | portable multimeter calibrator | portable multimeter calibrator | portable calibrator | portable calibrator] Original text/source link]