The fast Brinell hardness tester uses a far-infrared camera, a single-sided infrared dual point light source, and specialized analysis software to automatically measure Brinell hardness. It has the characteristics of automation, high recognition rate, applicability to any material surface, no material selection, not affected by daily visible light, and anti-interference ability; Simultaneously equipped with both manual and automatic modes; Advanced features such as self calibration mode
Rapid Brinell hardness testerKHT600M-QUICK
This machine implements the latest national standards:
GB/T231.2 The MetalclothRockwell hardness test 第Part 2: Hardness Testerinspection
How to translate
International StandardISO6506.2 “Metallic Materials-Brinell Hardness Test-Part2:Verification and Calibration of Testing Machine” American StandardASTME-10 “Test Method for Brine
ll Hardness of Metallic Materials”
lThe biggest feature of this model is its speed, automation, convenience, and accuracy. The measurement system can automatically output values by simply tapping the button. No need for manual intervention, dragging the mouse, or operating the keyboard.
lThe Brinell measurement system has fast speed, no material selection, and high recognition rate. Ordinary turning, milling, grinding of surfaces, sawing of cross-sections, surfaces filed by hand, surfaces produced by rough sanding wheels, rusted surfaces, and even damaged and incomplete indentations can all be automatically identified and measured. Because of the characteristics of not selecting raw materials and high recognition rate, it is particularly suitable for industrial practical sites.
l
The measurement system is both fixed and flexible: it can be fixed on the host to achieve integrated measurement; It can also be portable and handheld, enabling flexible and mobile measurement, serving multiple machines. It is also suitable for application scenarios where parts are large and difficult to move.Rapid Brinell hardness tester |
Experimental force:3000kgf (29400N)、1500Kgf (14700N)、1000Kgf (9800N)、750Kgf(7355N)、 500Kgf (4900N)、250Kgf (2452N)、187.5Kgf (1839N)、125Kgf (1226N)、100Kgf (980N)、 62.5Kgf(612.9N) |
Test force accuracy:00.5%Test force loading time: |
~8 seconds, in compliance with ASTM E10 standard |
Test force holding time: 2-99 seconds, adjustablePressure ball diameter:10mmThe5mm The |
2.5mm ASTME10 standard hard alloy ball |
Measurement range:( |
3.18~653)HBW Hardness indication accuracy: |
Hardness range( |
HBW) |
Indication error% |
Repeatability of indication% |
≤ 125 |
±3.0 |
≤3.0 |
125<HBW≤225 |
±2.5 225 |
≤2.5 |
> |
±2.0 200≤2.0 |
Maximum allowable height of the sample: 15mm |
Maximum distance from the center of the pressure head to the machine wall:0mmMaximum external dimensions |
:530mm×230mm×750mm |
Power supply: single-phase, AC,110V/220V,50~60Hz,4A |
Indentation measurement method:Automatic indentation reading systemNet weight |
|


|